Biomechanical analysis of ocular diseases and its in vitro study methods

Abstract

Ocular diseases are closely related to the physiological changes in the eye sphere and its contents. Using biomechanical methods to explore the relationship between the structure and function of ocular tissue is beneficial to reveal the pathological processes. Studying the pathogenesis of various ocular diseases will be helpful for the diagnosis and treatment of ocular diseases. We provide a critical review of recent biomechanical analysis of ocular diseases including glaucoma, high myopia, and diabetes. And try to summarize the research about the biomechanical changes in ocular tissues (e.g., optic nerve head, sclera, cornea, etc.) associated with those diseases. The methods of ocular biomechanics research in vitro in recent years are also reviewed, including the measurement of biomechanics by ophthalmic equipment, finite element modeling, and biomechanical analysis methods. And the preparation and application of microfluidic eye chips that emerged in recent years were summarized. It provides new inspiration and opportunity for the pathogenesis of eye diseases and personalized and precise treatment.

Introduction

The study of biomechanics can provide the mechanical forces information involved in biological processes from organs to cells level [1,2,3]. Each tissue inside and outside the eyeball has a certain tension, which is interrelated and affects each other in a closed spherical organ [4]. More and more attention has been paid to the study of the biomechanics of eye tissues to achieve the purpose of prevention and treatment of ocular diseases [5]. The pathogenesis of common and refractory eye disease has been studied through computational simulation modeling of eye tissue, to explore the biomechanical mechanism of ophthalmic diseases. It will lay a solid foundation for the development of modern ophthalmology precision medicine [6, 7].

At present, biomechanical studies on pathological ophthalmology-related diseases mainly focus on glaucoma [8], high myopia [9], and diabetic eye disease [10]. Biomechanics is an important part of the research on the pathogenesis, prevention, and treatment of glaucoma. High intraocular pressure (IOP) is the main risk factor for visual function injury. The increase of outflow resistance of the trabecular meshwork is the main factor in the increase of IOP. The increase of IOP will lead to iris deformation, pupil block, changes in the flow field of the anterior atrial, thinning of optic nerve fiber layer and lamina cribrosa (LC), visual function injury, etc. [11]. IOP is an important biomechanical factor directly related to glaucoma. Eye contents include aqueous humor, lens, and vitreous body, among which aqueous humor has the greatest influence on IOP. However, normal IOP can also lead to glaucoma, which is known as normal-tension glaucoma. In addition to the IOP, the biomechanical characteristics of the optic nerve head (ONH), LC, sclera, iris, pupillary block, aqueous humor outflow system, and trabecular meshwork associated with glaucoma also should be discussed [12, 13].

Moreover, the sclera is very important in the biomechanical study of myopia pathogenesis and treatment. Besides working with the cornea to maintain a refractive state, the sclera provides stable mechanical support to delicate ocular structures such as the retina and optic papilla. Diabetes causes many ocular complications, such as corneal epithelium, corneal stroma, corneal endothelium, which may result in further biomechanical changes in microstructure. Therefore, the biomechanical research progress related to the above diseases was summarized in the following text (Table 1).

Table 1 Biomechanical analysis of various ocular diseases

In vitro biomechanics research methods for ophthalmic diseases are mainly divided into medical equipment, finite element modeling, and the recently emerging microfluidic research [14]. Ophthalmic physiotherapy equipment like Corvis ST is mainly used to directly measure IOP or biomechanical parameters. The finite element modeling method is still focused on the study of glaucoma disease. Microfluidic-based eye chip research includes the preparation of eyeball drugs using on microreactor. The preparation of eye tissue chips through in vitro cell culture and microfluidic-based drug-release devices, etc., may provide a new idea for personalized eye disease research. Finite element modeling and in vitro eye organ chips may provide direction for the subsequent development and application of smart wearable contact lenses.

The main purpose of this review is to summarize the biomechanical research hotspots of common ophthalmic diseases in recent years, and to introduce the in vitro research methods around the biomechanics of these ocular diseases, hoping to provide reference for relevant researchers.

Biomechanical analysis of ocular diseases

Biomechanics is a powerful tool for the study of ophthalmic diseases. This part summarizes the research progress of common ophthalmic diseases in recent years, including glaucoma, high myopia and diabetic ocular diseases, hoping to help researchers have an understanding of this research area and provide new ideas for future research.

Glaucoma

The progressive death of retinal ganglion cells (RGC) in ONH is the main cause of glaucoma blindness. Poor aqueous humor drainage can lead to an increase in IOP, resulting in optic nerve papilla injury. Damage to the optic nerve caused by pathological structural changes can also lead to glaucoma even under normal IOP. This suggests that a full understanding of the biomechanical properties of the peripapillary sclera and the cornea may be helpful for the research and treatment of glaucoma [2].

It is necessary to develop a method for modeling individual differences in the optic nerve. Schwaner et al. [15] proposed a biomechanical modeling method of the rat optic nerve papilla for individual glaucoma studies. The three rat models of ONH are greater than that of human. The rat model of glaucoma ONH strain results were compared to learn more about the connection between the biomechanics of glaucoma and cell death. However, due to the model simplification, this method also has limitations. So, to refine the model, they [16] constructed and analyzed a model of the optic nerve head with an individual specific geometry, in which the sclera was modeled as a matrix reinforced by collagen fibers. The situation of elevated IOP was simulated. Through the data of rat glaucoma study, the mechanism of biomechanical influence on glaucoma retinal ganglion cell lesions in individual differences could be further explored. They also [17] established a parameterized model of the rat optic nerve for sensitivity studies. The results indicated that scleral properties had an important impact on the biomechanics of the rat optic nerve. Scleral thickness, scleral fiber arrangement, scleral fiber hardness, and scleral matrix hardness were the most influential parameters on the biomechanics of the optic nerve. So, for future modeling studies, specific values of these parameters should be determined to provide a more accurate research model of individual differentiation.

Moreover, glaucoma with normal intraocular pressure was also studied. Chen et al. [18] concluded that the optic nerve strain caused by axial extension is closely related to the unilateral normal-tension glaucoma pathological mechanism, similar to the cornea and sclera biomechanics. In patients with thin central cornea thickness, the correlation between central cornea thickness the strain rate of the optic nerve, and the strain rate of orbital fat were statistically significant [19]. Additionally, by measuring the response of the deformation of the sclera around the astrocyte layer and the adjacent papillary in mice to the increase of IOP, Korneva et al. [20] investigated the biomechanics of ONH and peripapillary sclera in a mouse model with glaucoma. The results showed that the mechanical behavior of the astrocyte layer and peripapillary sclera at the site of glaucoma neuron injury and remodeling changed dynamically over time.

Biomechanical mechanisms are thought to partly explain glaucoma optic neuropathy, Wei et al. [21] analyzed dynamic corneal response parameters by comparing normal-tension glaucoma subjects, hypertension glaucoma subjects, and control subjects. Results showed that the corneal deformation of normal-tension glaucoma was more pronounced than in hypertension glaucoma or the control group, people with hypertension glaucoma showed no significant difference in corneal deformation ability compared with the control group. The thinner the cornea, the lower the IOP, and the more easily the cornea deformation. These factors should be considered in the diagnosis of glaucoma.

High myopia

With the increase of myopia, excessive axial extension after the eyeball shows the biomechanics of stretching, followed by a series of retinopathy can cause significant vision loss. These may result in temporary or permanent loss of vision. The biomechanical changes of the sclera and cornea caused by myopia are the focus of myopia biomechanical research. To study the volume change of collagen fiber bundle structure behind sclera in highly myopic human eyes, Markov et al. [22] used the wide-angle X-ray scattering method to locate collagen in the sclera of non-myopic and highly myopic eyes. In high myopia, the normal post-scleral collagen microstructure changes greatly. These changes may reflect remodeling of the posterior sclera during axial elongation and/or mechanical adaptation to tissue pressure caused by fluid pressure or eye movement. Progression of myopia is thought to be associated with weakened scleral biomechanics, which results in ocular tissue deformation and axial elongation.

To strengthen weakened sclera and control myopia, scleral cross-linking has been proposed. The biomechanics of scleral weakening and scleral strengthening after cross-linking in myopia is not entirely clear. For investigating the effect of lysine oxidase (LOX) alone or in combination with Genipin on sclera cross-linking in myopic eyes, Wang et al. [23] compared the effects of cross-linking LOX and Genipin on the biomechanics and fixation index of guinea pig sclera. LOX crosslinked sclera in normal and myopic eyes. The cross-linking effect of LOX was weaker than that of Genipin, and the catalytic cross-linking effect of LOX and Genipin was not found in this study. Levy et al. [24] used Genipin to inhibit the scleral cyclic softening in tree shrews to investigate the increase in experimental myopia and scleral cross-linking. The results revealed that the sclera of young tree shrews was inelastic and cyclically softened by the cyclic tensile load.

It is well known that myopia alters the scleral structure and biomechanical properties, but its effect on corneal biomechanics is not well known. To further evaluate the corneal of myopia, Kang et al. [25] used an optical coherence tomography (OCT)-indentation probe, and found that the corneas of chicks with high myopia were more curved and softer on all IOP tests. Han et al. [26] explored that when there is a higher spherical equivalent in different degrees of myopia, the corneal stress–strain index is lower. It is suggested that the corneal mechanical strength might be compromised for high myopia. In addition, Yu et al. [27] observed the biomechanics and height of corneal after small incision lenticule extraction (SMILE) and laser-assisted subepithelial keratoplasty (LASEK). The effect of SMILE on corneal biomechanics might be smaller than that of LASEK in early postoperative removal of unit corneal tissue, but is comparable in long-term observation. Liu et al. [28] found that femtosecond-assisted LASEK (FS-LASIK) had less effect on corneal biomechanics than LASIK when assuming the same central corneal thickness for high myopia. The risk of corneal dilation after LASEK is lower than that after FS-LASIK.

Furthermore, highly myopic eyes are at significantly increased risk of many different secondary diseases due to morphological and structural changes [29]. High myopia may result in the risk of retinal detachment after lens surgery. Understanding the associated risk profile is clinically important. Grytz et al. [9] found that both myopia and glaucoma are chronic diseases leading to the connective tissue remodeling of the sclera and ONH. The mechanobiology behind connective tissue remodeling between the two diseases is essentially different, with different homeostasis control mechanisms.

Diabetic ocular disease

The blood glucose value increases in diabetes, which may result in a lot of eye diseases. It can affect the front and back segments of the eye and lead to severe vision defects and even blindness. The retina tissues could be acutely affected by diabetes, almost all ocular diseases may happen in diabetic patients [10]. The structure of the cornea and sclera will also be influenced by poor blood glucose control [7]. A series of studies will provide the possibility of simple and non-invasive treatment for diabetic patients, as well as may find new methods for early treatment of related complications.

Sahin et al. [30] measured IOP with an ocular response analyzer (ORA) and Goldman applanation tonometer (GAT) to explore the corneal biomechanical changes and their influence on IOP measurement in diabetic patients. The results showed that diabetes affected the corneal biomechanics and resulted in lower corneal hysteresis than the healthy controls, leading to clinical ocular hypertension. Kotecha et al. [31] used ORA to evaluate corneal hysteresis and corneal response in diabetic patients and nondiabetic patients, and explored that changes in corneal biomechanics in diabetic patients may be related to blood glucose concentration.

Furtherly, Scheler et al. [32] explored corneal hysteresis and corneal resistance factor in poorly controlled diabetes was more severe than those of healthy people and well-controlled diabetic subjects. It suggested that biomechanical properties of the cornea depending on glycemic control. Bao et al. [33] also found the mechanical stiffness of the cornea of diabetic rabbits was significantly increased, which was manifested by the increase in corneal thickness and tangent modulus. Ramm et al. [34] detected corneal hysteresis and found corneal resistance factor was significantly increased in diabetic patients by using ORA and Corvis ST. And then, Ramm et al. [35] collected and evaluated age, IOP, and central corneal thickness of diabetic patients and healthy people. They assessed the effect of disease-specific factors and established a reliable sensitivity and specificity for diabetes mellitus values.

Same findings as above, Beato et al. [36] compared the corneal biomechanics after phacoemulsification in diabetes people and those without diabetes. The recovery of corneal hysteresis in diabetes was slower than in nondiabetic patients, while both groups were subjected to a significant and continuous decrease of the IOP and corneal resistance factor in phacoemulsification. In addition, Terai et al. [37] found that ONH stiffness was notably increased in diabetic rats compared with the peripapillary sclera, which may be related to non-enzymatic collagen cross-linking mediated by late glycation end products induced by diabetic hyperglycemia. To make clear whether these biomechanical changes have adverse risk factors for IOP regulation in diabetic people with glaucoma, further studies are needed to do.

In vitro biomechanical study methods of eye tissue

Noninvasive in vitro study can provide more accurate and a large amount data for ocular biomechanics analysis, this part summarizes the recent study about the biomechanical measurement equipment, the finite element modeling and microfluidic chip in vitro method for ocular disease study.

Ocular biomechanical analysis using medical ophthalmic equipment

Measurement of ocular parameters with ophthalmic professional medical equipment is an important basis for clinical diagnosis and treatment [38]. In the study of human eye data, a clinical visual corneal biomechanics analyzer (Corvis ST) is mainly used. Glaucoma, for example, to assess the reaction parameters on the same day the dynamics of the cornea patients, springback tonometer measurement of IOP, corneal thickness, and ocular axial length. It is beneficial for the analysis of normal-tension glaucoma and unilateral normal-tension glaucoma pathogenic factors [39]. Hong et al. [40] reported a new biomechanical parameter corneal flattening speed by Corvis ST measurement to determine whether it is related to the diagnosis of glaucoma with normal IOP. The speed of corneal flattening inward in normal IOP glaucoma was smaller than that in normal IOP glaucoma, which was statistically significant.

Additionally, Pradhan et al. [41, 42] compared corneal biomechanics in pseudoexfoliation syndrome, sham stripping pseudoexfoliation glaucoma, primary open-angle glaucoma, and healthy controls with Corvis ST, and found the nearly the same results for those corneal biomechanical parameters. Vinciguerra et al. [43] compared Corvis ST and Goldmann tonometers in primary and normal open-angle glaucoma, high IOP, and control group. Moreover, they compared the dynamic corneal response parameters of Corvis ST for primary high IOP, the high IOP, and the control group, suggesting that corneal biomechanics may be an important factor affecting IOP measurement. Abnormal corneal biomechanics and its significant correlation with field parameters may be harmful reasons for the generation of open-angle glaucoma with normal IOP.

Besides, Eliasy et al. [44] considered the biomechanical behavior of healthy cornea measured by Corvis ST tension measurement using the stress–strain index. The method can help optimize corneal surgery such as refractive surgery and the introduction of corneal implants. Perez-rico et al. [45] studied the influence of diabetes on Corvis ST measurement process. Comparisons were made using the ocular response analyzer and Corvis ST analysis to determine the influence of disease duration, hyperglycemia, and hemoglobin levels on these parameters. Poor glycemic control in diabetics affects corneal biomechanics as measured by the eye response analyzer and Corvis ST, which may suggest that high IOP is independent of central corneal thickness. The measurement of corneal biomechanics should be considered in clinical applications.

Ocular tissue biomechanical analysis using finite element method

Finite element analysis software is favored by researchers because of its powerful simulation function (Table 2). A variety of complex structures can be established, the database contains the rich unit type and material properties to facilitate the establishment of the physical model. The construction of the physical model of a reusable can exert various load cases and the model can be visualized structure of internal stress and strain concentration conditions. This method can effectively avoid animal experiments and medical ethics problems. The combination of computer simulation data and clinical study results provides a possible solution for the prediction and treatment of early ocular abnormalities. Due to its advantages, researchers often apply finite element analysis to the study of mechanic-related pathogenesis and treatment of glaucoma and eye surgery.

Table 2 Ocular tissue biomechanical finite element analysis

Ocular finite element modeling

Finite element (FE) modeling is generally regarded as an effective method to quantitatively analyze the pathogenesis of glaucoma. In recent years, research has focused on building local eye models. Karimi et al. [46] simulated the biomechanical stress and strain of ONH using FE method, and various mesh elements were used to investigate the effect of mesh type on the results. The 20-node hexahedron element produces the most accurate results in complex models. The calculation results of the 10-node tetrahedral element and 20-node hexahedral element are very similar and can be exchanged for a short calculation time. Linear element types do not produce acceptable results. To get closer to the real eye tissue, Zhou et al. [47] proposed a new biomechanical material model of the whole eye. Collagen content in the eye tissue measured by X-ray scattering was expressed by Zernike polynomials covering the cornea and sclera. Based on the user-defined material model, a fine mesh FE model with the specific geometry of the human eye was established. The model was then used in iterative inverse modeling studies to derive material parameters.

However, local eye tissue modeling responses are often not comprehensive enough to include global eye structure models, so more effective information may be obtained in the study of ophthalmic disease biomechanics. As increased IOP may be a major danger factor for glaucoma, Dai et al. [48] constructed a global FE eye model and simulated the effect of elevated IOP on eye structure. A refined global eye model was established using ANSYS software to study the relationship between IOP and biomechanical response. First, the pressure transfer process of elevated IOP was analyzed to simulate the effect of elevated IOP on glaucoma. Then, biomechanical responses of the anterior eye segment under different pressures were analyzed by simulating non-adhesion of the iris and posterior sclera. The model not only simulates the effects of elevated IOP on ocular structure, but also reveals the process of pressure transfer from the anterior segment to the ONH. The local mechanical properties of the eye structure obtained from the global model are consistent with the previous results. This global model may shed light on multifactorial glaucoma research. Studies have shown that eyeball design can affect the results of biomechanical analysis. Issarti et al. [49] assessed the mechanical contribution of intraocular structures to corneal deformation by establishing the whole eyeball FE mode. The corneal mechanical deformation under different mechanical conditions revealed that the lens, iris, and muscle are also the main factors that should be considered.

Glaucoma-related ocular tissue biomechanical finite element analysis

Biomechanical properties of ONH and lamina cribrosa (LC)

At present, biomechanical analysis of glaucoma tissues by finite element method is the main application type to explore the pathogenic mechanism and the application of treatment. ONH cell damage is widely recognized as the direct cause of blindness in glaucoma, but the pathway of its occurrence remains unclear.

Glaucoma is partly characterized by elevated and fluctuating IOP, which in turn loads the head of the optic nerve. In addition, tissue viscoelasticity strongly affects the mechanical response of ONH to mechanical load, but its viscoelastic mechanical properties remain unclear. To determine these properties, Safa et al. [50] made micromechanical tests and constructed a mixed model containing two phase materials of the viscoelastic solid matrix. It was proved that the viscoelastic mechanical response of ONH could be revealed by the mechanism of fluid flow and solid matrix viscoelasticity. Understanding these parameters will facilitate the establishment of in vitro research models and experiments, and further explain the pathogenesis of glaucoma under different conditions.

In recent years, ONH injury caused by other factors is also the focus of researchers. Wang et al. [51] discovered that the hardness of Bruch’s membrane-choroid complex is equal to or higher than that of other eye tissues, it may have a non-negligible influence on the ONH deformation induced by high IOP. Similarly, Feola et al. [52] explored how the anatomy and swelling of choroids affect the ONH by developing finite element models of ONH. Results showed that choroids may have a great influence on the biomechanics of ONH. In addition, Jin et al. [53] developed a FE model of a normal eye, and predicted ocular pulse amplitude, choroidal expansion could affect the biomechanics of ONH during the cardiac cycle. Ma et al. [54] investigated the correlation between IOP-induced local displacement of the ONH and the thickness of the peripapillary sclera, the results suggest that the greater backward movement of ONH relative to the surrounding tissues, the thinner peripapillary tissue, and also the LC may play a major role in preventing excessive backward displacement of ONH during acute elevated IOP.

Kim et al. [55] explored the effects of heart rate changes on the dynamic biomechanical characteristic of ocular pulse and ONH. Ocular pulse amplitude OPA, pulse volume, and ONH deformation decreased with the increase of AR heart rate, while LC became hard. The influence of changes in blood pressure/heart rate on ONH sclerosis may have important implications for the pathology of glaucoma. FE simulation was used by Shin et al. [56], results revealed that mechanical stress and strain concentrated in the ONH region by the abduction of the optic nerve sheath were much greater than that of the elevated IOP. This supports the new concept that glaucoma optic neuropathy may be caused at least in part by external traction of the optic nerve, and not just the pressure from the inside of the eye on the optic nerve.

The pore shape and size of the LC are predictors of the mechanical damage to the optic nerve tissue in glaucoma. To determine the deformation of the neural tissue in the LC pore, Voorhees et al. [57] established computational models of LC, with different nonlinear anisotropy and neural microstructures based on the tissue sections of sheep’s eyes. The microscopic structure of the LC results in localized mechanical changes in neural tissue. In addition, they further measured the significant damage to the LC nerve tissue [58]. Similarly, Karimi et al. [59] proposed FE models of three human eye posterior poles, including LC microstructures and distributed nerve tissue composed of retinal axons. These models were used to estimate the stress and strain of the LC and dispersed neural tissue under acute elevated IOP.

Sclera, iris, pupil obstruction, aqueous outflow system, trabecular meshwork

Schwaner et al. [60] studied the biomechanics of rat sclera by inverse FE model. The scleral surface displacement was first measured by digital image correlation. The sclera was modeled as a nonlinear material embedded with collagen fibers, and used a differential evolution algorithm to fit the model displacement to the experimental data. The material properties identified were beneficial for the study of glaucoma. Pant et al. [61] carried out an image-based reverse modeling analysis to quantitatively explore the iris properties in angle-closure glaucoma people.

The iris in glaucomatous patients was stiffer than that of healthy people. It may provide a reference for the study of the mechanism in angle-closure glaucoma. Wang et al. [62] constructed 21 eye finite element models to simulate the influence of the pupillary blocking force with various pupil diameters and iris–lens channel distance. It was shown that the influence of the iris–lens channel on pupillary blocking force is more notable than pupil diameter. It would be the main potential risk for primary angle-closure glaucoma.

It is difficult to obtain the accurate location and hydrodynamic data of aqueous outflow resistance with increased IOP. Zhang et al. [63] found that the change of trabecular meshwork permeability has the greatest influence on hydromechanical parameters of trabecular meshwork and Schlemm’s canal. By using the bidirectional fluid–structure coupling simulation method, Wang et al. [64] confirmed the hardness of the trabecular meshwork in glaucoma patients is higher, and outflow from both normal eyes and glaucoma appears to be associated with stiffness of the trabecular meshwork combining finite element modeling and ocular imaging device. It also provides a reference for further research on the factors regulating the biomechanical properties of the trabecular meshwork.

Finite element analysis of cornea and other eye tissues

Evaluation of corneal biomechanics is the basis for the study of ocular surgery and the accuracy of IOP measurement. The biomechanical responses of the cornea and adipose tissue in dynamic IOP tests were evaluated using the inverse finite element method and inverse model [65]. Pandolfi et al. [66] have taken into account the biological tissues with multiple physical properties and typical characteristics of the ocular environment, and custom geometric models are constructed based on the parameters of imaging and in vivo testing. The acquired patient-specific model can provide information on the post-operative shape of the cornea [67].

Montanino et al. [68] proposed a numerical model for non-contact testing that could evaluate the mechanical parameters of the human cornea. This test consists of a fast injection of air on the corneal anterior surface and concluded that the internal fluid is vitally important for simulation. The effect of different scleral stiffness on corneal biomechanical deformation response under airbag load was described by the finite element model. Nquyen et al. [69] conducted the FE study of the biomechanical effect of sclera on corneal deformation response. Inflatable deformation in a two-dimensional axisymmetric fixed FE model of the whole eye is generally considered to be entirely due to IOP and corneal characteristics. The current study shows that the harder the sclera, the greater the limitation of corneal deformation, independent of IOP. It may have important clinical application value to explain cornea response under aeration load under pathological conditions. Qin et al. [70] proposed a way to measure corneal elastic modulus based on Corvis ST results. Based on the calculated elastic modulus, the corneal apical displacement was simulated by the finite element method, and the simulated corneal apical displacement was consistent with the experimental results. Based on the relationship between force and displacement of shallow spherical shells, the method of determining corneal elastic modulus based on the Corvis test is simple and effective.

Moreover, Karimi et al. [71] studied the biomechanics of healthy cornea and corneal conus by combining clinical data, FE, and artificial neural network, and established a new biomechanics-based keratoconus eye diagnosis method. Clinical biomechanical parameters of healthy and only keratoconus could be obtained by non-contact tonometer. According to the corneal geometry, the finite element model of each cornea and the same boundary and loading conditions were applied not only to confirm the biomechanical FE parameters, but also to calculate the amount of von Mises stress at the apex of the cornea. Then clinical biomechanical data and von Mises stress were combined into the artificial neural network algorithm to distinguish healthy cornea from keratoconus based on the resulting von Mises stress. These findings not only have important implications for eye care professionals to identify keratoconus as an important clinical and surgical tool, but also provide quantitative and accurate methods for understanding the biomechanical properties of keratoconus.

Additionally, Zhang et al. [72] introduced a new way to measure the mechanical stiffness of healthy cornea and corneal cone. FE modeling performed the analysis of healthy cornea and corneal cone. Stress–strain index was proposed and evaluated in earlier studies as a parameter, and was used as a method for corneal stiffness study. SSI maps can estimate regional variations in biomechanical stiffness of the cornea surface. Rahmati et al. [73] used coupled FE optimization algorithm to estimate the viscoelastic properties of healthy corneas and Keratoconus corneas. This algorithm is a non-invasive technique that can accurately estimate the viscoelastic cornea properties.

Corneal surgery mechanical impact study has established the 3D FE human eye model to verify the laser in situ corneal grinding with surgery (LASIK) on corneal biomechanics behavior, Bao et al. [74] found that when considering LASIK surgery of corneal biomechanical behavior, improving the match of clinical measurements and LASIK predictions. This result has important implications for the development of planning tools for refractive surgery. Mohamed et al. [75] developed a new numerical simulation method for corneal transplantation to study the stress and deformation of corneal tissue and donor transplantation during corneal endothelial transplantation. The results clearly showed that bubble pressure is vitally important for the stress and strain of the cornea and the stiffness and thickness of the corneal membrane.

In arcuate keratectomy, Truffer et al. [76] proposed an arcuate keratotomy scheme for personalized FE simulation. Virtual surgery was performed on patients, and a numerical model was used to optimize the surgical parameters of arcuate keratotomy, which could improve the reliability of postoperative astigmatism and reduce the risk of overcorrection. Knaus et al. [77] investigated the effect of ciliary muscle contraction on lens adjustment by using a 3D FE model. The results showed that the ciliary muscle section had a synergistic effect: the circular section contributed the most to the increase of lens thickness, while the longitudinal and radial sections did the opposite. Conversely, the function of the longitudinal and radial parts is to shift the lens forward, as opposed to the circular part. The finite element model demonstrates the complex interaction of ciliary muscle three segments during lens deformation and transformation during adjustment.

In addition, due to car accidents, earthquakes, gang fights, and other reasons, glass fragments can collide with the eye, causing numerous scars, which can permanently affect vision. Complications from collisions with the eye and subsequent damage to various parts of the eye can be difficult to diagnose. Karimi et al. [78] constructed a 3D human eye model to evaluate the impact of glass fragments on human eyes. As the speed of glass fragments increases, the stress in the center of the vestibule, like the cornea, water body, and iris, increases. But for parts located outside/behind the eye, especially the optic nerve, a reduction in pressure is observed by increasing speed, with little damage to the optic nerve. These findings not only help to understand the stress/tension in the eye at different speeds, but also help to provide ophthalmologists with initial information for better diagnosis after eye injuries from fragment-like (small object impact) injuries.

To develop a detailed human orbital FE model and verify it by analyzing its behavior under blunt trauma pressure. Foletti et al. [79] used a modified 3D FE human head, which is the most complex model developed to date, to advance the understanding of the mechanics of optic nerve injury. Singman et al. [80] developed a model of the head with a biofiber-mesh eye socket. This study represented the first published biological neural simulation using the full length of the optic nerve, in which the eye socket is integrated with the entire head. Song et al. [81] used a 3D FE model to study the orbital bone development of congenital microphthalmia. Periocular biomechanics mainly focuses on the medial center of the orbital wall and gradually extends to the entire orbital wall. Deck et al. [82] study the FE eye model of retinal hemorrhage in shaken baby syndrome. In the retina, the slightest tremor causes four times as much pressure as the most severe shock, and shaking a baby causes extreme eye strain compared to falling.

Microfluidic method in ocular chip

Numerical simulation has been recognized as an effective means of eye disease biomechanical research, but for most studies, the eye model is often simplified, and biomechanics analysis was conducted under ideal conditions without considering other factors, which may lead to the results deviation of the numerical analysis and the actual situation [107]. In recent years, researchers have shown great interest in alternatives to in vitro models of the eye (Table 3), allowing for the addition of dynamic fluid flows to better mimic the physiological structure of the eye [108, 109].

Table 3 In vitro biomechanical study methods of eye tissue

Organ microfluidic chip is a transformative technology that can reproduce the human organs in vitro disease model [110]. The development of these technologies will increase our understanding of the different structure of eyes of basic physiology, enables us to examine the unknown aspects of eye disease pathogenesis, and as a substitute for clinically relevant to assess eye treatment [111].

In addition, since tears include many biomarkers, contact lens sensors can be used to directly measure different parameters such as glucose, urea, protein concentrations, ions [112], IOP, and corneal temperature, noninvasively. Microelectromechanical machining technology [113] enables contact lenses with detection electrodes and tiny structures to be used as wearable devices for biomarkers monitoring and delivery of drugs for treating eye diseases [114]. The combination of eyeball physiological information monitoring and drug delivery system will become the trend of personalized ophthalmic disease diagnosis and treatment [115].

Ophthalmic disease research platform based on microfluidic chip

Biomarkers contained in tears provide an important reference for the non-invasive understanding of physiological disease processes. Karns et al. [83] developed a microfluidic homogeneous immunoassay chip for a rapid, quantitative and specific measurement of endogenous tear protein (lactoferrin) biomarkers in human tears. The accuracy of the immunoassay chip was within 15% of ELISA, and the detection limit was 3 ± 2 nM (Fig. 1a). Sonobe et al. [84] proposed a novel paper-based ELISA analytical device for assessing tear lactoferrin in dry eye people. Results showed that lactoferritin concentration was correlated with Schirmer test values and tear film rupture time, but negatively correlated with ocular surface disease index, luciferin, and rose Bengal scores.

Fig. 1
figure 1

a Microfluidic tear component analysis platform proposed by Karns et al. [83] Reproduced with permission from Ref. [83]. Copyright 2011, American Chemical Society. b Colorimetric μPad device for tear electrolyte analysis proposed by Yetisen et al. [87]. Reproduced with permission from Ref. [87]. Copyright 2020, Royal Society of Chemistry. c Wearable intraocular pressure sensor and detection system based on contact lens proposed by Araci et al. [90] Reproduced with permission from Ref. [90]. Copyright 2019, Royal Society of Chemistry

Microfluidic fluorescence analyzer has a broad application prospect as a preclinical and clinical diagnostic instrument. Commercial fluorescence microscopes, however, are expensive and time-consuming. It is not conducive to the portable and wearable applications of microfluidic chips. Pestana et al. [85] developed a dedicated low-cost fluorescent microfluidic device reader for reading interlayer immunofluorescence analysis equipment and for detecting vascular endothelial growth factor ligand concentrations in eye fluid samples. In addition, the detection of tears facilitates early diagnosis of eye diseases, and monitoring of high-risk subjects. Yetisen et al. [86] designed a μPad system for tear electrolyte analysis (Na+, K+, Ca2+, and pH), including a sample collection capillary, a sample dilution reservoir, and a μPad for electrolyte analysis, and its fluorescence output was measured using a smartphone readout device.

Changes in the composition of tears are signals of ocular and systemic metabolic processes, so they can be utilized to assess physical health. Moreddu et al. [87] proposed paper microfluidics integrated into laser-implanted commercial contact lenses for detection of biomarkers. In vitro measurements include the colorimetric way, which collects, stores, and analyzes readings using a customized tear diagnostic smartphone app prototype (Fig. 1b). The application demonstrates the device’s potential for discrete measurements during medical diagnosis. In addition, discovery further developed a laser-cut wearable contact lens sensor for the analytes monitoring in tears [88]. Microfluidic systems are implanted in commercial contact lenses via CO2 laser ablation. The microchannel consists of a central ring consisting of four branches, and the biosensor is embedded in a microcavity at the end of the branch. Colorimetric readout based on the nearest neighbor model is carried out by MatLab algorithm of smart phone. They then further developed a laboratory platform for contact lenses using multiaxial femtosecond laser ablation [89] to quickly and accurately etch microfluidic networks on the surface of contact lenses. Production of functional microfluidics components such as flow valves, resistors, multi-entry geometry, and dispensers were realized with custom seven-axis femtosecond laser systems.

The measurement of IOP is also an important means of detecting glaucoma. The expansion strain sensor, which works by detecting volume changes in microfluidic channels, is highly sensitive to biaxial strain and is made only from soft and transparent materials, making it easy to integrate with smartphones. These characteristics have advantages for contact lens-based IOP sensing applications. Because in the measurement of IOP, the noise generated by eye movement and pulsation will affect the measurement of IOP signal, and the use of electronic components for filtering will bring inconvenience to the human body. The microfluidic equivalent circuit can realize analog/digital conversion, filtering, and other logical operations through flow channel design and fluid parameter modification. Araci et al. [90] adopted the microfluidic equivalent circuit to achieve low stability and noise suppression of wearable microfluidic sensors and improve the signal–noise ratio of ophthalmic applications. Suppression of noise caused by eye pulsation, blinking, etc., without the need for electronic components was realized (Fig. 1c).

In addition, the study of biomechanics and pharmacokinetics of eyeballs by constructing in vitro eyeball model and using the microfluidic method is also one of the research hotspots in recent years. Beissner et al. [91] applied a previously designed dynamic microtissue engineering system with a pre-validated human corneal structure to obtain an improved test platform. The platform offers a huge opportunity to improve common in vitro drug testing procedures.

The cost and ethical criticism of ocular drug development of animal pharmacokinetic studies made it necessary to develop an in vitro model to study corneal. Bennet et al. [92] prepared a porous membrane embedded in the microfluidic platform to separate the chip into top and base sides. Immortalized human corneal epithelial cells were grown on the membrane to create a microengineered corneal epithelial chip (Corneal chip), a model that can imitate the environment of the human cornea. The current model in the further study of angiogenesis in vitro lack of extensibility (Fig. 2a). Ko et al. [93] developed a plastic-based microfluidic chip for in vitro reconstruction of the three-dimensional vascular network, and carried out the human eye angiogenesis model on the injection molding microfluidic chip. The chip provides a simple fluid mode for constructing cell culture microenvironments. This model can be used not only in normal and pathological vascular studies, but also in basic studies of ocular neovascularization.

Fig. 2
figure 2

a An in vitro eyeball model platform developed by a microfluidic eyeball cell chip was used to study pharmacokinetics proposed by Bennet et al. [92]. Reproduced with permission from Ref. [92]. Copyright 2018, Royal Society of Chemistry. b The human corneal barrier and blink reconstruction platform based on the microfluidic chip is used to study the development of ophthalmic drugs proposed by Abdalkader et al. [95] Reproduced with permission from Ref. [95]. Copyright 2020, Royal Society of Chemistry

Isopropanol silicone oil is commonly used as a tamponade to treat complex retinal detachment and proliferative vitreoretinopathy, which is prone to emulsification in vivo and may cause inflammation. Lu et al. [94] used a microfluidic device and an in vitro “eye-on-a-chip” to evaluate whether the addition of polymer silicone oil could facilitate the injection and removal of silicone oil. The results show that the presence of polymer silicone oil can increase the tensile viscosity and prevent the decomposition of silicone oil from the matrix to form emulsion droplets.

Human corneal epithelium coexists with tear fluid and exhibits its barrier function under dynamic blinking conditions. However, the current in vitro culture environment of corneal epithelial cells lacks dynamic flow conditions to reproduce the shearing stress during blinking. Abdalkader et al. [95] developed a microfluidic platform that can dynamically cultivate the human corneal barrier by recreating blinking. It is expected that the multi-corneal barrier device on a chip will open new possibilities for the development of ophthalmic drugs and will contribute to the study of the effects of blink shear stress on the ocular surface (Fig. 2b). Bai et al. [96] conducted the preparation of a novel 3D chip cornea using primary mouse corneal epithelial and endothelial cells. The separation and culture scheme of primary corneal epithelial cells and endothelial cells was used to construct a microfluidic 3D microengineered cornea based on primary cells. This chip overcomes the shortcomings of 2D cell culture and realizes corneal function and delivery of drugs.

Preparation of ophthalmic drug and drug delivery device based on the microfluidic chip

For ophthalmic diseases, surgical implants and implanted drugs are currently the main treatment mode. However, from the perspective of patient comfort (such as regular insertion and removal of implants), they are not accepted by users. Currently, the main non-invasive treatment options, such as eye drops, are only 1–3% of the drug can reach the tissues in the eye. Improving the controlled release of ophthalmic drugs is a key challenge in addressing the local administration of hormones and drugs.

Polymer particles are capable of drug loading and controlled release, and their biocompatibility and biodegradability make them powerful tools for non-invasive drugs. The glaucoma drug betaprolol, for example, is released more slowly and for longer when combined with polymer ion-exchange resin particles suspended in an adhesive medium. However, the preparation of uniform particle size, controllable particle size, and high drug loading, good degradation of polymer particles is the main problem. The preparation of polymer particles based on a microfluidic chip provides the possibility for the preparation of polymer particles. Polylactic acid–glycolic acid copolymer (PLGA) is one of the most widely used biodegradable materials [116]. Leon et al. [97] used microfluidic technique to prepare multidrug polymer particles for the treatment of external glaucoma. They manufactured multidrug-loaded biodegradable polymer particles by loading Latanoprost and dexamethasone onto monodisperse microparticles (approximately 150 μm in diameter) of biodegradable PLGA, the monodisperse oil-in-water emulsion was generated using a capillary microfluidic device, which was then evaporated by the thin film to produce monodisperse particles with a diameter of = 150 um and a standard deviation of  < 5%. The parameters of drug loading, drug release, in vitro adhesion, and the measurement of local drug use were verified and optimized. The study is expected to reduce the number of times patients receive eye drops and improve the effectiveness of drug delivery.

In addition to polymer micron particles, drug nanosuspensions were found to have high bioavailability. Drug nanosuspensions are composed of micron drug particles suspended in a dispersed medium and stabilized by polymers or surfactants. The current preparation of nanosuspension drugs is mainly by grinding large drug particles or by precipitation of drug molecules to construct nanodrugs, and the main development direction is the latter. Ali et al. [98] compared the preparation of ophthalmic hydrocortisone nanosuspension by the microfluidic nanoprecipitation method and wet grinding method. Hydrocortisone is a widely used steroid to treat all kinds of eye inflammation. The particle size, shape, and crystallinity of the suspensions prepared by the two methods were characterized. The results showed that hydrocortisone nanosuspension was developed by the microfluidic nanoprecipitation method. As a novel, simple and economical drug nanocrystalization technology, the drug action time of nanometer suspensions could be significantly prolonged.

So far, researchers have developed many eyes organ in vitro models, from 2D cell culture models based on single-cell culture, cell culture models, or trained models, to the three-dimensional organs, 3D printing chip systems, and organs, organs including local and overall, in vitro drug screening and disease research contribution. In particular, the combination of microfluidic technology can make the in vitro study of drug delivery effects very useful [99]. Micro-drug delivery devices are also vitally important in the controlled release of drugs. Precision ophthalmic drug delivery devices combine mechanical, electronic, and microfluidic functions [100].

Since most ophthalmic drugs were delivered via topical eye drops, there is a huge demand for the controlled release of drugs from ocular biomaterials. Kaczmarek et al. [101] constructed a long-term wear-resistant microfluid hydrogel contact lens by adjusting the composition of biomaterials to realize the controlled release of dexamethasone over 60 days. Phan et al. [102] explored a new in vitro eye model for fluconazole release from a variety of commercial contact lenses. The eye model is prepared by 3D printing and filled with PDMS, which can obtain a cheap model of the eye and eyelid. Use injection pump as a fluid driving force. Compared with 3D printed eye models, the vials had higher drug release. The drug is quickly released from the contact lens within the first 2 h, followed by a plateau. Rapid drug release can be achieved by using vials as a drug release system. The volume of tears significantly speeds up the drug release process. Subsequently, they further refined the model [103] of the eyeball and eyelid masses to simulate physiological tear volume. The integration of human or animal corneas or human corneas will allow for more complex in vitro eye studies. Combined with the injection pump, the platform was constructed for studies to evaluate drug delivery and deposition in contact lenses.

Similarly, because most used ophthalmology in vitro drug release study is carried out under static conditions, and are not fully considering the influence of flow dynamics, need to take into account the tear volume and flow in the body, Pimenta et al. [104] designed a microfluidic unit to simulate the tear continuous, the volumetric flow rate, and its small volume. The release kinetics of a drug system containing diclofenac was compared with that under static condition by using the microfluidic chip, and the results showed that the release kinetics under dynamic condition was slower. Subsequently, they investigated [105] plasma-assisted grafting of 2-acrylamide-2-methylpropane sulfonic acid or 2-(methylacryloxy) ethyl dimethyl-(3-sulfopropyl ammonium hydroxide) to the surface modification of hydrophilic acrylic materials for the preparation of intraocular lenses. Its load of the endophthalmitis antibiotic moxifloxacin was shown to significantly prolong the drug release time (up to 12 days) in vitro. Silva et al. [106] used moxifloxacin hydrochloride imprinted silicon-based hydrogel as a contact lens material for prolonged drug release. Contact lenses can be used as a platform for drug release for the treatment of eye infections, but their typical release period still needs to be extended. The microfluidic unit was used to simulate the flow of ocular surface fluid. The results showed that the release period of Staphylococcus aureus and Staphylococcus epidermidis was about 2 weeks.

Conclusion and future trend

Ocular diseases are closely related to the physiological changes in the eyeball and its contents. Using biomechanical methods to explore the relationship between the structure and function of eye tissue at multiple scales is helpful to reveal the pathological processes. Therefore, this text summarized the biomechanical research status of several ophthalmic diseases (including glaucoma, high myopia, and diabetic eye disease). The biomechanical research of eyeballs based on computer modeling is the mainstream at present, but the simplified model and ideal conditions may result in different conclusions from the real situation. The combination with ophthalmic medical equipment may be helpful to the diagnosis and surgical treatment of eye diseases. With the development of microfluidic organ chips and wearable microfluidic chips, more and more refined biological models of the eyeball can be realized through 3D cell culture, which undoubtedly plays a good supplement to the in vitro biomechanical research of the eyeball. In addition, the preparation of eyeball nanomedicine based on microfluidic chips, wearable intraocular metabolite sensors, and eyeball drug controlled-release microfluidic devices provide new ideas for the treatment of systemic ophthalmic diseases. It is believed that more theoretical biomechanical models of eyeball diseases will be combined with in vitro physiological models in the subsequent studies to promote the progress of ophthalmic medicine.

Availability of data and materials

Not applicable.

Abbreviations

IOP:

Intraocular pressure

ONH:

Optic nerve head

LC:

Lamina cribrosa

RGC:

Retinal ganglion cells

LOX:

Lysine oxidase

SMILE:

Small incision lenticule extraction

LASEK:

Laser-assisted subepithelial keratoplasty

ORA:

Ocular response analyzer

GAT:

Goldman applanation tonometer

FE:

Finite element

PLGA:

Polylactic acid–glycolic acid copolymer

References

  1. Khong EWC, Chan HHL, Watson SL, Lim LL. Pregnancy and the eye. Curr Opin Ophthalmol. 2021;32:527–35.

    Article  Google Scholar 

  2. Park SA, Komaromy AM. Biomechanics of the optic nerve head and sclera in canine glaucoma: a brief review. Vet Ophthalmol. 2021;24:316–25.

    Article  Google Scholar 

  3. Komaromy AM, Koehl KL, Park SA. Looking into the future: gene and cell therapies for glaucoma. Vet Ophthalmol. 2021;24:16–33.

    Article  Google Scholar 

  4. Hancock SE, Wan CR, Fisher NE, Andino RV, Ciulla TA. Biomechanics of suprachoroidal drug delivery: from benchtop to clinical investigation in ocular therapies. Expert Opin Drug Del. 2021;18:777–88.

    Article  Google Scholar 

  5. Ferrara M, Lugano G, Sandinha MT, Kearns VR, Geraghty B, Steel DHW. Biomechanical properties of retina and choroid: a comprehensive review of techniques and translational relevance. Eye. 2021;35:1818–32.

    Article  Google Scholar 

  6. Bujor IA, Iancu RC, Istrate SL, Ungureanu E, Iancu G. Corneal biomechanical changes in third trimester of pregnancy. Medicina-Lithuania. 2021. https://0-doi-org.brum.beds.ac.uk/10.3390/medicina57060600.

    Article  Google Scholar 

  7. Wang XY, Xu GH, Wang W, Wang JJ, Chen LL, He MG, Chen ZL. Changes in corneal biomechanics in patients with diabetes mellitus: a systematic review and meta-analysis. Acta Diabetol. 2020;57:973–81.

    Article  Google Scholar 

  8. Sng CCA, Ang M, Barton K. Central corneal thickness in glaucoma. Curr Opin Ophthalmol. 2017;28:120–6.

    Article  Google Scholar 

  9. Grytz R, Yang HL, Hua Y, Samuels BC, Sigal IA. Connective tissue remodeling in myopia and its potential role in increasing risk of glaucoma. Curr Opin Biomed Eng. 2020;15:40–50.

    Article  Google Scholar 

  10. del Buey MA, Casas P, Caramello C, Lopez N, de la Rica M, Subiron AB, Lanchares E, Huerva V, Grzybowski A, Ascaso FJ. An update on corneal biomechanics and architecture in diabetes. J Ophthalmol. 2019;2019:7645352.

    Google Scholar 

  11. Boote C, Sigal IA, Grytz R, Hua Y, Nguyen TD, Girard MJA. Scleral structure and biomechanics. Prog Retin Eye Res. 2020. https://0-doi-org.brum.beds.ac.uk/10.1016/j.preteyeres.2019.100773.

    Article  Google Scholar 

  12. Kling S, Hafezi F. Corneal biomechanics—a review. Ophthal Physiol Opt. 2017;37:240–52.

    Article  Google Scholar 

  13. Ma JN, Wang Y, Wei PH, Jhanji V. Biomechanics and structure of the cornea: implications and association with corneal disorders. Surv Ophthalmol. 2018;63:851–61.

    Article  Google Scholar 

  14. Tseng RC, Chen CC, Hsu SM, Chuang HS. Contact-lens biosensors. Sensors. 2018. https://0-doi-org.brum.beds.ac.uk/10.3390/s18082651.

    Article  Google Scholar 

  15. Schwaner SA, Kight AM, Perry RN, Pazos M, Yang HL, Johnson EC, Morrison JC, Burgoyne CF, Ethier CR. A methodology for individual-specific modeling of rat optic nerve head biomechanics in glaucoma. J Biomech Eng. 2018. https://0-doi-org.brum.beds.ac.uk/10.1115/1.4039998.

    Article  Google Scholar 

  16. Schwaner SA, Perry RN, Kight AM, Winder E, Yang HL, Morrison JC, Burgoyne CF, Ethier CR. Individual-specific modeling of rat optic nerve head biomechanics in glaucoma. J Biomech Eng. 2021. https://0-doi-org.brum.beds.ac.uk/10.1115/1.4049157.

    Article  Google Scholar 

  17. Schwaner SA, Feola AJ, Ethier CR. Factors affecting optic nerve head biomechanics in a rat model of glaucoma. J R Soc Interface. 2020. https://0-doi-org.brum.beds.ac.uk/10.1098/rsif.2019.0695.

    Article  Google Scholar 

  18. Hocaoglu M, Kara C, Sen EM, Ozturk F. Relationships between corneal biomechanics and the structural and functional parameters of glaucoma damage. Arq Bras Oftalmol. 2020;83:132–40.

    Article  Google Scholar 

  19. Aksoy NO, Cakir B, Aksoy YE, Boncukcu KD, Ozmen S, Celik E, Alagoz G. Effects of glaucoma and central corneal thickness on optic nerve head biomechanics. Inter Ophthalmol. 2021;41:1283–9.

    Article  Google Scholar 

  20. Korneva A, Kimball EC, Jefferys JL, Quigley HA, Nguyen TD. Biomechanics of the optic nerve head and peripapillary sclera in a mouse model of glaucoma. J R Soc Interface. 2020. https://0-doi-org.brum.beds.ac.uk/10.1098/rsif.2020.0708.

    Article  Google Scholar 

  21. Wei YH, Cai Y, Choy BNK, Li BB, Li RS, Xing C, Wang X, Tian T, Fang Y, Li M, Pan YZ. Comparison of corneal biomechanics among primary open-angle glaucoma with normal tension or hypertension and controls. Med J-Peking. 2021;134:1087–92.

    Google Scholar 

  22. Markov PP, Eliasy A, Pijanka JK, Htoon HM, Paterson NG, Sorensen T, Elsheikh A, Girard MJA, Boote C. Bulk changes in posterior scleral collagen microstructure in human high myopia. Mol Vis. 2018;24:818–33.

    Google Scholar 

  23. Wang M, Yang ZK, Li RQ, Fan H, Liu Y. Comparison of the biomechanics and fixation index of crosslinking between lysyl oxidase and genipin on guinea pig sclera, an animal model of defocus-induced high myopia. P Est Acad Sci. 2019;68:214–21.

    Article  Google Scholar 

  24. Levy AM, Fazio MA, Grytz R. Experimental myopia increases and scleral crosslinking using genipin inhibits cyclic softening in the tree shrew sclera. Ophthal Physiol Opt. 2018;38:246–56.

    Article  Google Scholar 

  25. Kang BS, Wang LK, Zheng YP, Guggenheim JA, Stell WK, Kee CS. High myopia induced by form deprivation is associated with altered corneal biomechanical properties in chicks. PLoS ONE. 2018. https://0-doi-org.brum.beds.ac.uk/10.1371/journal.pone.0207189.

    Article  Google Scholar 

  26. Han F, Li MD, Wei PH, Ma JN, Jhanji V, Wang Y. Effect of biomechanical properties on myopia: a study of new corneal biomechanical parameters. BMC Ophthalmol. 2020. https://0-doi-org.brum.beds.ac.uk/10.1186/s12886-020-01729-x.

    Article  Google Scholar 

  27. Yu MR, Chen MJ, Dai JH. Comparison of the posterior corneal elevation and biomechanics after SMILE and LASEK for myopia: a short- and long-term observation. Graef Arch Clin Exp. 2019;257:601–6.

    Article  Google Scholar 

  28. Liu MN, Shi WY, Liu X, Li N, Chen T, Gao H. Postoperative corneal biomechanics and influencing factors during femtosecond-assisted laser in situ keratomileusis (FS-LASIK) and laser-assisted subepithelial keratomileusis (LASEK) for high myopia. Laser Med Sci. 2021;36:1709–17.

    Article  Google Scholar 

  29. Ziemssen F, Lagreze W, Voykov B. Secondary diseases in high myopia. Ophthalmologe. 2017;114:30–43.

    Article  Google Scholar 

  30. Sahin A, Bayer A, Ozge G, Mumcuoglu T. Corneal biomechanical changes in diabetes mellitus and their influence on intraocular pressure measurements. Invest Ophth Vis Sci. 2009;50:4597–604.

    Article  Google Scholar 

  31. Kotecha A, Oddone F, Sinapis C, Elsheikh A, Sinapis D, Sinapis A, Garway-Heath DF. Corneal biomechanical characteristics in patients with diabetes mellitus. J Cataract Refr Surg. 2010;36:1822–8.

    Article  Google Scholar 

  32. Scheler A, Spoerl E, Boehm AG. Effect of diabetes mellitus on corneal biomechanics and measurement of intraocular pressure. Acta Ophthalmol. 2012;90:e447–51.

    Article  Google Scholar 

  33. Bao FJ, Deng ML, Zheng XB, Li LN, Zhao YP, Cao S, Yu AY, Wang QM, Huang JH, Elsheikh A. Effects of diabetes mellitus on biomechanical properties of the rabbit cornea. Exp Eye Res. 2017;161:82–8.

    Article  Google Scholar 

  34. Ramm L, Herber R, Spoerl E, Pillunat LE, Terai N. Measurement of corneal biomechanical properties in diabetes mellitus using the ocular response analyzer and the Corvis ST. Cornea. 2019;38:595–9.

    Article  Google Scholar 

  35. Ramm L, Herber R, Spoerl E, Pillunat LE, Terai N. Factors influencing corneal biomechanics in diabetes mellitus. Cornea. 2020;39:552–7.

    Article  Google Scholar 

  36. Beato JN, Esteves-Leandro J, Reis D, Falcao M, Rosas V, Carneiro A, Reis FF. Corneal biomechanical changes after uneventful phacoemulsification in patients with type 2 diabetes mellitus and patients without diabetes. Cornea. 2020;39:1073–9.

    Article  Google Scholar 

  37. Terai N, Spoerl E, Haustein M, Hornykewycz K, Haentzschel J, Pillunat LE. Diabetes mellitus affects biomechanical properties of the optic nerve head in the rat. Ophthalmic Res. 2012;47:189–94.

    Article  Google Scholar 

  38. Wolffsohn JS, Kollbaum PS, Berntsen DA, Atchison DA, Benavente A, Bradley A, Buckhurst H, Collins M, Fujikado T, Hiraoka T, et al. IMI—clinical myopia control trials and instrumentation report. Invest Ophth Vis Sci. 2019;60:M132–60.

    Article  Google Scholar 

  39. Chen YY, Wang TH, Huang JY, Su CC. Relationship of axial length and corneal biomechanical properties with susceptibility to unilateral normal-tension glaucoma. Graefes Arch Clin Exp Ophthalmol. 2022;260:255–64.

    Article  Google Scholar 

  40. Hong K, Wong IYH, Singh K, Chang RT. Corneal biomechanics using a scheimpflug-based noncontact device in normal-tension glaucoma and healthy controls. Asia-Pac J Ophthalmo. 2019;8:22–9.

    Google Scholar 

  41. Pradhan ZS, Deshmukh S, Dixit S, Gudetti P, Devi S, Webers CAB, Rao HL. A comparison of the corneal biomechanics in pseudoexfoliation syndrome, pseudoexfoliation glaucoma, and healthy controls using Corvis (R) scheimpflug technology. Indian J Ophthalmol. 2020;68:787–92.

    Article  Google Scholar 

  42. Pradhan ZS, Deshmukh S, Dixit S, Sreenivasaiah S, Shroff S, Devi S, Webers CAB, Rao HL. A comparison of the corneal biomechanics in pseudoexfoliation glaucoma, primary open-angle glaucoma and healthy controls using Corvis ST. PLoS ONE. 2020. https://0-doi-org.brum.beds.ac.uk/10.1371/journal.pone.0241296.

    Article  Google Scholar 

  43. Vinciguerra R, Rehman S, Vallabh NA, Batterbury M, Czanner G, Choudhary A, Cheeseman R, Elsheikh A, Willoughby CE. Corneal biomechanics and biomechanically corrected intraocular pressure in primary open-angle glaucoma, ocular hypertension and controls. British J Ophthalmol. 2020;104:121–6.

    Article  Google Scholar 

  44. Eliasy A, Chen KJ, Vinciguerra R, Lopes BT, Abass A, Vinciguerra P, Ambrosio R, Roberts CJ, Elsheikh A. Determination of corneal biomechanical behavior in-vivo for healthy eyes using CorVis ST tonometry: stress–strain index. Front Bioeng Biotech. 2019. https://0-doi-org.brum.beds.ac.uk/10.3389/fbioe.2019.00105.

    Article  Google Scholar 

  45. Perez-Rico C, Gutierrez-Ortiz C, Gonzalez-Mesa A, Zandueta AM, Moreno-Salgueiro A, Germain F. Effect of diabetes mellitus on corvis ST measurement process. Acta Ophthalmol. 2015;93:e193–8.

    Article  Google Scholar 

  46. Karimi A, Grytz R, Rahmati SM, Girkin CA, Downs JC. Analysis of the effects of finite element type within a 3D biomechanical model of a human optic nerve head and posterior pole. Comput Meth Prog Bio. 2021;198:105794.

    Article  Google Scholar 

  47. Zhou D, Abass A, Eliasy A, Studer HP, Movchan A, Movchan N, Elsheikh A. Microstructure-based numerical simulation of the mechanical behaviour of ocular tissue. J R Soc Interface. 2019. https://0-doi-org.brum.beds.ac.uk/10.1098/rsif.2018.0685.

    Article  Google Scholar 

  48. Dai PS, Zhao YL, Sheng HW, Li L, Wu J, Han H. Simulating the effect of elevated intraocular pressure on ocular structures using a global finite element model of the human eye. J Mech Med Biol. 2017. https://0-doi-org.brum.beds.ac.uk/10.1142/S0219519417500385.

    Article  Google Scholar 

  49. Issarti I, Koppen C, Rozema JJ. Influence of the eye globe design on biomechanical analysis. Comput Biol Med. 2021;135:104612.

    Article  Google Scholar 

  50. Safa BN, Read AT, Ethier CR. Assessment of the viscoelastic mechanical properties of the porcine optic nerve head using micromechanical testing and finite element modeling. Acta Biomate. 2021;134:379–87.

    Article  Google Scholar 

  51. Wang XF, Teoh CKG, Chan ASY, Thangarajoo S, Jonas JB, Girard MJA. Biomechanical properties of bruch’s membrane-choroid complex and their influence on optic nerve head biomechanics. Invest Ophth Vis Sci. 2018;59:2808–17.

    Article  Google Scholar 

  52. Feola AJ, Nelson ES, Myers J, Ethier CR, Samuels BC. The impact of choroidal swelling on optic nerve head deformation. Invest Ophth Vis Sci. 2018;59:4172–81.

    Article  Google Scholar 

  53. Jin YJ, Wang XF, Zhang L, Jonas JB, Aung T, Schmetterer L, Girard MJA. Modeling the origin of the ocular pulse and its impact on the optic nerve head. Invest Ophth Vis Sci. 2018;59:3997–4010.

    Article  Google Scholar 

  54. Ma YH, Kwok S, Sun JJ, Pan XL, Pavlatos E, Clayson K, Hazen N, Liu J. IOP-induced regional displacements in the optic nerve head and correlation with peripapillary sclera thickness. Exper Eye Res. 2020;200:108202.

    Article  Google Scholar 

  55. Jin YJ, Wang XF, Irnadiastputri SFR, Mohan RE, Aung T, Perera SA, Boote C, Jonas JB, Schmetterer L, Girard MJA. Effect of changing heart rate on the ocular pulse and dynamic biomechanical behavior of the optic nerve head. Invest Ophth Vis Sci. 2020;61:27.

    Article  Google Scholar 

  56. Shin A, Yoo L, Park J, Demer JL. Finite element biomechanics of optic nerve sheath traction in adduction. J Biomech Eng. 2017. https://0-doi-org.brum.beds.ac.uk/10.1115/1.4037562.

    Article  Google Scholar 

  57. Voorhees AP, Jan NJ, Austin ME, Flanagan JG, Sivak JM, Bilonick RA, Sigal IA. Lamina cribrosa pore shape and size as predictors of neural tissue mechanical insult. Invest Ophth Vis Sci. 2017;58:5336–46.

    Article  Google Scholar 

  58. Voorhees AP, Jan NJ, Sigal IA. Effects of collagen microstructure and material properties on the deformation of the neural tissues of the lamina cribrosa. Acta Biomater. 2017;58:278–90.

    Article  Google Scholar 

  59. Karimi A, Rahmati SM, Grytz RG, Girkin CA, Downs JC. Modeling the biomechanics of the lamina cribrosa microstructure in the human eye. Acta Biomater. 2021;134:357–78.

    Article  Google Scholar 

  60. Schwaner SA, Hannon BG, Feola AJ, Ethier CR. Biomechanical properties of the rat sclera obtained with inverse finite element modeling. Biomech Model Mechan. 2020;19:2195–212.

    Article  Google Scholar 

  61. Pant AD, Gogte P, Pathak-Ray V, Dorairaj SK, Amini R. Increased Iris stiffness in patients with a history of angle-closure glaucoma: an image-based inverse modeling analysis. Invest Ophth Vis Sci. 2018;59:4134–42.

    Article  Google Scholar 

  62. Wang WJ, Song HF, Liu ZC. Computational study on the biomechanics of pupil block phenomenon. Biomed Res Inter. 2019;2019:4820167.

    Google Scholar 

  63. Zhang J, Qian XQ, Zhang HX, Liu ZC. Fluid-structure interaction simulation of aqueous outflow system in response to juxtacanalicular meshwork permeability changes with a two-way coupled method. Cmes-Comp Model Eng. 2018;116:301–14.

    Google Scholar 

  64. Wang K, Johnstone MA, Xin C, Song SZ, Padilla S, Vranka JA, Acott TS, Zhou K, Schwaner SA, Wang RKK, et al. Estimating human trabecular meshwork stiffness by numerical modeling and advanced OCT imaging. Invest Ophth Vis Sci. 2017;58:4809–17.

    Article  Google Scholar 

  65. Jannesari M, Kadkhodaei M, Mosaddegh P, Kasprzak H, Behrouz MJ. Assessment of corneal and fatty tissues biomechanical response in dynamic tonometry tests by using inverse models. Acta Bioeng Biomech. 2018;20:39–48.

    Google Scholar 

  66. Pandolfi A. Cornea modelling. Eye Vis. 2020. https://0-doi-org.brum.beds.ac.uk/10.1186/s40662-019-0166-x.

    Article  Google Scholar 

  67. Maklad O, Eliasy A, Chen KJ, Theofilis V, Elsheikh A. Simulation of air puff tonometry test using Arbitrary Lagrangian-Eulerian (ALE) deforming mesh for corneal material characterisation. Int J Env Res Pub He. 2020;17:54.

    Article  Google Scholar 

  68. Montanino A, Angelillo M, Pandolfi A. A 3D fluid-solid interaction model of the air puff test in the human cornea. J Mech Behav Biomed. 2019;94:22–31.

    Article  Google Scholar 

  69. Nguyen BA, Roberts CJ, Reilly MA. Biomechanical impact of the sclera on corneal deformation response to an air-puff: a finite-element study. Front Bioeng Biotech. 2019;6:210.

    Article  Google Scholar 

  70. Qin X, Tian L, Zhang HX, Chen XY, Li L. Evaluation of corneal elastic modulus based on corneal visualization scheimpflug technology. Biomed Eng Online. 2019;18:42.

    Article  Google Scholar 

  71. Karimi A, Meimani N, Razaghi R, Rahmati SM, Jadidi K, Rostami M. Biomechanics of the healthy and keratoconic corneas: a combination of the clinical data, finite element analysis, and artificial neural network. Curr Pharm Design. 2018;24:4474–83.

    Article  Google Scholar 

  72. Zhang HX, Eliasy A, Lopes B, Abass A, Vinciguerra R, Vinciguerra P, Ambrosio R, Roberts CJ, Elsheikh A. Stress-strain index map: a new way to represent corneal material stiffness. Front Bioeng Biotech. 2021. https://0-doi-org.brum.beds.ac.uk/10.3389/fbioe.2021.640434.

    Article  Google Scholar 

  73. Rahmati SM, Razaghi R, Karimi A. Biomechanics of the keratoconic cornea: theory, segmentation, pressure distribution, and coupled FE-optimization algorithm. J Mech Behav Biomed. 2021. https://0-doi-org.brum.beds.ac.uk/10.1016/j.jmbbm.2020.104155.

    Article  Google Scholar 

  74. Bao FJ, Wang JJ, Cao S, Liao N, Shu B, Zhao YP, Li YY, Zheng XB, Huang JH, Chen SH, et al. Development and clinical verification of numerical simulation for laser in situ keratomileusis. J Mech Behav Biomed. 2018;83:126–34.

    Article  Google Scholar 

  75. Mohamed S, Coccarelli A, Mauro A, Massarotti N, Romano MR, Romano V, Nithiarasu P. A novel numerical modelling approach for keratoplasty eye procedure. Biomech Model Mechan. 2019;18:1429–42.

    Article  Google Scholar 

  76. Truffer O, Abler D, Pajic B, Grabner G, Kraker H, Buchler P. Optimization of surgical parameters based on patient-specific models: application to arcuate keratotomy. J Cataract Refr Surg. 2019;45:1084–91.

    Article  Google Scholar 

  77. Knaus KR, Hipsley A, Blemker SS. The action of ciliary muscle contraction on accommodation of the lens explored with a 3D model. Biomech Model Mech. 2021;20:879–94.

    Article  Google Scholar 

  78. Karimi A, Razaghi R, Biglari H, Sera T, Kudo S. Collision of the glass shards with the eye: a computational fluid-structure interaction model. J Chem Neuroanat. 2018;90:80–6.

    Article  Google Scholar 

  79. Foletti JM, Martinez V, Graillon N, Godio-Raboutet Y, Thollon L, Guyot L. Development and validation of an optimized finite element model of the human orbit. J Stomatol Oral Maxi. 2019;120:16–20.

    Google Scholar 

  80. Li Y, Singman E, McCulley T, Wu CW, Daphalapurkar N. The biomechanics of indirect traumatic optic neuropathy using a computational head model with a biofidelic orbit. Front Neurol. 2020;11:346.

    Article  Google Scholar 

  81. Song DY, Zhang JW, Yuan BW, Rong QG, Li DM. A finite element analysis model is suitable for biomechanical analysis of orbital development. J Craniofac Surg. 2021;32:2546–50.

    Article  Google Scholar 

  82. Nadarasa J, Deck C, Meyer F, Bourdet N, Raul JS, Willinger R. Development of a finite-element eye model to investigate retinal hemorrhages in shaken baby syndrome. Biomech Model Mechan. 2018;17:517–30.

    Article  Google Scholar 

  83. Karns K, Herr AE. Human tear protein analysis enabled by an alkaline microfluidic homogeneous immunoassay. Anal Chem. 2011;83:8115–22.

    Article  Google Scholar 

  84. Sonobe H, Ogawa Y, Yamada K, Shimizu E, Uchino Y, Kamoi M, Saijo Y, Yamane M, Citterio D, Suzuki K, Tsubota K. A novel and innovative paper-based analytical device for assessing tear lactoferrin of dry eye patients. Ocul Surf. 2019;17:160–6.

    Article  Google Scholar 

  85. Pestana N, Walsh D, Hatch A, Hahn P, Jaffe GJ, Murthy SK, Niedre M. A dedicated low-cost fluorescence microfluidic device reader for point-of-care ocular diagnostics. J Med Devices. 2013. https://0-doi-org.brum.beds.ac.uk/10.1115/1.4023995.

    Article  Google Scholar 

  86. Yetisen AK, Jiang N, Tamayol A, Ruiz-Esparza GU, Zhang YS, Medina-Pando S, Gupta A, Wolffsohn JS, Butt H, Khademhosseini A, Yun SH. Paper-based microfluidic system for tear electrolyte analysis. Lab Chip. 2017;17:1137–48.

    Article  Google Scholar 

  87. Moreddu R, Elsherif M, Adams H, Moschou D, Cordeiro MF, Wolffsohn JS, Vigolo D, Butt H, Cooper JM, Yetisen AK. Integration of paper microfluidic sensors into contact lenses for tear fluid analysis. Lab Chip. 2020;20:3970–9.

    Article  Google Scholar 

  88. Moreddu R, Wolffsohn JS, Vigolo D, Yetisen AK. Laser-inscribed contact lens sensors for the detection of analytes in the tear fluid. Sensor Actuat B-Chem. 2020;317:128183.

    Article  Google Scholar 

  89. Moreddu R, Nasrollahi V, Kassanos P, Dimov S, Vigolo D, Yetisen AK. Lab-on-a-contact lens platforms fabricated by multi-axis femtosecond laser ablation. Small. 2021. https://0-doi-org.brum.beds.ac.uk/10.1002/smll.202102008.

    Article  Google Scholar 

  90. Araci IE, Agaoglu S, Lee JY, Yepes LR, Diep P, Martini M, Schmidt A. Flow stabilization in wearable microfluidic sensors enables noise suppression. Lab Chip. 2019;19:3899–908.

    Article  Google Scholar 

  91. Beissner N, Mattern K, Dietzel A, Reichl S. DynaMiTES—a dynamic cell culture platform for in vitro drug testing Part 2-ocular DynaMiTES for drug absorption studies of the anterior eye. Eur J Pharm Biopharm. 2018;126:166–76.

    Article  Google Scholar 

  92. Bennet D, Esttack Z, Reid T, Kim J. A microengineered human corneal epithelium-on-a-chip for eye drops mass transport evaluation. Lab Chip. 2018;18:1539–51.

    Article  Google Scholar 

  93. Ko J, Lee Y, Lee S, Lee SR, Jeon NL. Human ocular angiogenesis-inspired vascular models on an injection-molded microfluidic chip. Adv Healthc Mater. 2019. https://0-doi-org.brum.beds.ac.uk/10.1002/adhm.201900328.

    Article  Google Scholar 

  94. Lu YJ, Chan YK, Lau LH, Chao YC, Shih KC, Lai SM, Wong D, Shum HC. Adhesion of silicone oil and emulsification: an in vitro assessment using a microfluidic device and ‘Eye-on-a-Chip.’ Acta Ophthalmol. 2019;97:313–8.

    Article  Google Scholar 

  95. Abdalkader R, Kamei K. Multi-corneal barrier-on-a-chip to recapitulate eye blinking shear stress forces. Lab Chip. 2020;20:1410–7.

    Article  Google Scholar 

  96. Bai J, Fu HJ, Bazinet L, Birsner AE, D’Amato RJ. A method for developing novel 3d cornea-on-a-chip using primary murine corneal epithelial and endothelial cells. Front Pharmacol. 2020;11:453.

    Article  Google Scholar 

  97. Leon RAL, Somasundar A, Badruddoza AZM, Khan SA. Microfluidic fabrication of multi-drug-loaded polymeric microparticles for topical glaucoma therapy. Part Syst Char. 2015;32:567–72.

    Article  Google Scholar 

  98. Ali HSM, York P, Ali AMA, Blagden N. Hydrocortisone nanosuspensions for ophthalmic delivery: a comparative study between microfluidic nanoprecipitation and wet milling. J Control Release. 2011;149:175–81.

    Article  Google Scholar 

  99. Kutlehria S, Singh M. Role of in vitro models for development of ophthalmic delivery systems. Crit Rev Ther Drug. 2021;38:1–31.

    Article  Google Scholar 

  100. Zhang HB, Jackson JK, Chiao M. Microfabricated drug delivery devices: design, fabrication, and applications. Adv Funct Mater. 2017;27:1703606.

    Article  Google Scholar 

  101. Kaczmarek JC, Tieppo A, White CJ, Byrne ME. Adjusting biomaterial composition to achieve controlled multiple-day release of dexamethasone from an extended-wear silicone hydrogel contact lens. J Biomat Sci-Polym E. 2014;25:88–100.

    Article  Google Scholar 

  102. Phan CM, Bajgrowicz M, Gao HY, Subbaraman LN, Jones LW. Release of fluconazole from contact lenses using a novel in vitro eye model. Optometry Vision Sci. 2016;93:387–94.

    Article  Google Scholar 

  103. Phan CM, Walther H, Gao HY, Rossy J, Subbaraman LN, Jones L. Development of an in vitro ocular platform to test contact lenses. Jove-J Vis Exp. 2016. https://0-doi-org.brum.beds.ac.uk/10.3791/53907.

    Article  Google Scholar 

  104. Pimenta AFR, Valente A, Pereira JMC, Pereira JCF, Filipe HP, Mata JLG, Colaco R, Saramago B, Serro AP. Simulation of the hydrodynamic conditions of the eye to better reproduce the drug release from hydrogel contact lenses: experiments and modeling. Drug Deliv Transl Re. 2016;6:755–62.

    Article  Google Scholar 

  105. Pimenta AFR, Vieira AP, Colaco R, Saramago B, Gil MH, Coimbra P, Alves P, Bozukova D, Correia TR, Correia IJ, et al. Controlled release of moxifloxacin from intraocular lenses modified by Ar plasma-assisted grafting with AMPS or SBMA: an in vitro study. Colloid Surface B. 2017;156:95–103.

    Article  Google Scholar 

  106. Silva D, de Sousa HC, Gil MH, Santos LF, Oom MS, Alvarez-Lorenzo C, Saramago B, Serro AP. Moxifloxacin-imprinted silicone-based hydrogels as contact lens materials for extended drug release. Eur J Pharm Sci. 2021;156:105591.

    Article  Google Scholar 

  107. Boutin ME, Hampton C, Quinn R, Ferrer M, Song MJ. 3D engineering of ocular tissues for disease modeling and drug testing. In: Bharti K, editor. Pluripotent stem cells in eye disease therapy. Cham: Springer; 2019. p. 171–93.

    Chapter  Google Scholar 

  108. Prina E, Mistry P, Sidney LE, Yang J, Wildman RD, Bertolin M, Breda C, Ferrari B, Barbaro V, Hopkinson A, et al. 3D Microfabricated scaffolds and microfluidic devices for ocular surface replacement: a review. Stem Cell Rev Rep. 2017;13:430–41.

    Article  Google Scholar 

  109. Jahagirdar D, Bangde P, Jain R, Dandekar P. Degenerative disease-on-a-chip: developing microfluidic models for rapid availability of newer therapies. Biotechnol J. 2021. https://0-doi-org.brum.beds.ac.uk/10.1002/biot.202100154.

    Article  Google Scholar 

  110. Shi HH, Nie KX, Dong B, Long MQ, Xu H, Liu ZC. Recent progress of microfluidic reactors for biomedical applications. Chem Eng J. 2019;361:635–50.

    Article  Google Scholar 

  111. Manafi N, Shokri F, Achberger K, Hirayama M, Mohammadi MH, Noorizadeh F, Hong JX, Liebau S, Tsuji T, Quinn PMJ, Mashaghi A. Organoids and organ chips in ophthalmology. Ocul Surf. 2021;19:1–15.

    Article  Google Scholar 

  112. Shi HH, Jiang SF, Liu B, Liu ZC, Reis NM. Modern microfluidic approaches for determination of ions. Microchem J. 2021. https://0-doi-org.brum.beds.ac.uk/10.1016/j.microc.2021.106845.

    Article  Google Scholar 

  113. Shi HH, Nie KX, Dong B, Long MQ, Liu ZC. Automatic in situ synthesis system for polypeptide biochip based on microfluidic mixer. IEEE T Nanobiosci. 2021;20:116–25.

    Article  Google Scholar 

  114. Shi HH, Zeng YN, Zhou YN, Zhang WWH, CX, Zhao YL, Chen Z. Wearable tesla valve-based sweat collection device for sweat colorimetric analysis. Talanta. 2022;240:123208.

    Article  Google Scholar 

  115. Savariraj AD, Salih A, Alam F, Elsherif M, AlQattan B, Khan AA, Yetisen AK, Butt H. Ophthalmic sensors and drug delivery. Acs Sensors. 2021;6:2046–76.

    Article  Google Scholar 

  116. Su Y, Zhang BL, Sun RW, Liu WF, Zhu QB, Zhang X, Wang RR, Chen CP. PLGA-based biodegradable microspheres in drug delivery: recent advances in research and application. Drug Deliv. 2021;28:1397–418.

    Article  Google Scholar 

Download references

Acknowledgements

This work was supported financially by the National Natural Science Foundation of China (Grant No. 62101231), the Hunan Provincial Natural Science Foundation, China (Grant No. 2021JJ40618), and the Jiangxi Provincial Natural Science Foundation, China (Grant No. 20202BABL212003). We would like to acknowledge the participants who donated their time and effort for the study.

Funding

This work was supported financially by the National Natural Science Foundation of China (Grant No. 62101231), the Hunan Provincial Natural Science Foundation, China (Grant No. 2021JJ40618), and the Jiangxi Provincial Natural Science Foundation, China (Grant No. 20202BABL212003).

Author information

Authors and Affiliations

Authors

Contributions

YLZ wrote the main text and completed the draft manuscript. GHH and YWY performed the literature and data collections. ZW and XHL completed tables and figures making and content proofreading. HHS proposed the concept and revised the manuscript. All authors read and approved the final manuscript.

Corresponding authors

Correspondence to Yali Zhao or Huanhuan Shi.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

All authors consent to the publication.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhao, Y., Hu, G., Yan, Y. et al. Biomechanical analysis of ocular diseases and its in vitro study methods. BioMed Eng OnLine 21, 49 (2022). https://0-doi-org.brum.beds.ac.uk/10.1186/s12938-022-01019-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s12938-022-01019-1

Keywords