Fig. 2

From: Investigation of safety for electrochemotherapy and irreversible electroporation ablation therapies in patients with cardiac pacemakers

Fig. 2

The interference of a single electroporation pulse (a, c, e) and a sequence of four electroporation pulses (b, d, f) with 1000 V amplitude, 100 μs duration, and 5 kHz repetition rate. Electroporation pulses were delivered before (a, b), during (c, d) or after (e, f) the ventricular pacing pulse in the medium with conductivity of 1.57 S/m. For comparison with the effects observed in the media of lower conductivity at 1000 V and 3000 V amplitudes see Additional file 1: Figures S1 and S2