Skip to main content

Objective assessment of spasticity by pendulum test: a systematic review on methods of implementation and outcome measures

Abstract

Background

Instrumented pendulum test is an objective and repeatable biomechanical method of assessment for spasticity. However, multitude of sensor technologies and plenty of suggested outcome measures, confuse those interested in implementing this method in practice. Lack of a standard agreement on the definition of experimental setup and outcome measures adds to this ambiguity and causes the results of one study not to be directly attainable by a group that uses a different setup. In this systematic review of studies, we aim to reduce the confusion by providing pros and cons of the available choices, and also by standardizing the definitions.

Methods

A literature search was conducted for the period of 1950 to the end of 2019 on PubMed, Science Direct, Google Scholar and IEEE explore; with keywords of “pendulum test” and “Spasticity”.

Results

Twenty-eight studies with instrumented pendulum test for assessment of spasticity met the inclusion criteria. All the suggested methods of implementation were compared and advantages and disadvantages were provided for each sensor technology. An exhaustive list categorized outcome measures in three groups of angle-based, angular velocity-based, and angular acceleration-based measures with all different names and definitions.

Conclusions

With the aim of providing standardized methodology with replicable and comparable results, sources of dissimilarity and ambiguity among research strategies were found and explained with the help of graphical representation of pendulum movement stages and corresponding parameters on the angular waveforms. We hope using the provided tables simplify the choices when implementing pendulum test for spasticity evaluation, improve the consistency when reporting the results, and disambiguate inconsistency in the literature.

Background

Spasticity according to European Thematic Network to Develop Standardized Measures of Spasticity is ‘‘disordered sensory-motor control, resulting from an upper motor neuron lesion, presenting as intermittent or sustained involuntary activation of muscles’’ [1]. This complex motor disorder which is caused by diverse sources of upper motor neuron lesions, affects patients with stroke [2], spinal cord injury [3], traumatic brain injury [4], multiple sclerosis [5] and cerebral palsy [6]. Spasticity can create severe pain and hence may negatively affect the quality of life. Spasticity can interfere with movement and can lead to stiff, painful joints [7, 8]. This disorder affects more than 12 million persons in the world and interferes with the natural movements of the patients [9, 10]. Billions of dollars are spent in pharmaceutical industry to develop anti-spasticity drugs, though lack of repeatable and objective outcome measures hinders the success [8]. It is generally agreed that spasticity is easy to recognize, but not so easy to quantify. Quantitative assessment of spasticity and spasms is crucial in evaluating treatment interventions. Although quantitative clinical scales have been proposed and are currently in use [11,12,13,14], but they lack objectivity. This means that with lower resolution and repeatability, which are inherent with subjective measures, perceived level of spasticity or its improvement/worsening can contain errors or small changes be unrecognized.

To overcome objectivity issue, one of the promising methods that was introduced about 70 years ago was pendulum test [15]. However, such objective quantification systems that can be easily used in clinics are still missing; although care should be taken that processing techniques may also affect numerical outcomes of any test [16]. Many attempts can be found in the literature that aimed at providing instrumented pendulum test of spasticity [17,18,19] and at times for stiffness [20]. Still, when it comes to methods of implementation and choice of outcome measures, the literature is very diverse and at times non-consistent. In other words, with each group of researchers having their own methodology, although the level of spasticity can be relatively assessed in each experiment data, comparing the results between experiments are very difficult. This makes practical use of the published research difficult, particularly for those who want to actually implement this method. Although Review studies can be found on pendulum test of spasticity, they have mainly focused on psychometric properties of this method [21, 22]. Although these reports state that pendulum test method is reliable and valid, they have not focused on the compatibility of results between researchers with different methodologies. In this article we have exhaustively searched the published scientific literature to come up with a clear list of options for practical methods of implementation. We have further tried to find the sources of incompatibility in the reported results and particularly considered effect of available choices of outcome measures for pendulum test of spasticity.

Results

Description of studies

As summarized in Fig. 1, initial search provided 1581 articles. After removing duplicates, 1554 articles remained from which 1456 were excluded because of not being eligible by examining the title and abstract. From the remaining 98 articles, 70 were removed based on full text assessment for eligibility. Three more relevant studies were found through the articles’ references. Therefore, 28 articles met the eligibility criteria and were included in the review (Table 3).

Fig. 1
figure 1

Flow chart of the search and selection process of the systematic review of objective assessment of spasticity by pendulum test for methods of implementation and outcome measures

Assessment of spasticity

Methods of assessment of spasticity can be broadly categorized into clinical, biomechanical, and electrophysiological [23] methods. Clinical methods, that are gold standards in clinic routines, are semi-qualitative. Although clinical scales have been proposed to provide quantitative outcomes, they suffer from lack of objectivity. This means, there is usually inter- and intra-clinician differences while evaluating levels of spasticity; and the resolution is often lower than when using a fully objective device. Biomechanical and electrophysiological methods involve sensors or instruments and can provide objective outcome measures.

A handful of clinical measures such as (CSI: composite spasticity index [24], HAT: hypertonia assessment tool [25], and TSS: triple spasticity scale [26], etc.) have been proposed and used in clinical assessments. However, the most routine method of assessment of spasticity by clinicians are the following four scales that are compared in detail in Table 1, Table 2: Ashworth Scale (AS) [11], Modified Ashworth Scale (MAS) [12], Tardieu Scale (TS) [13], and Modified Tardieu Scale (MTS) [14].

Table 1 Comparison of Ashworth and Modified Ashworth scales
Table 2 Comparison of Tardieu and Modified Tardieu scales

All these scales assess spasticity by manual estimation of increased resistance of specific muscle groups to passive motion. From historical point of view, in 1954 Tardieu et al. introduced a 6-point scale to assess spasticity for the first time which was speed dependent (explained in the following paragraphs).

A decade after that and independently, in 1964, Ashworth proposed a 5-point scale which did not take the velocity dependence of spasticity into account. This scale was modified in 1987 by Bohannon and Smith to deal with the accumulation of most of the scores towards the lower end of the scale. They performed this by including an extra category (1+ , Table 1), making it a 6-point scale, and also modified the definitions slightly. In 1999, Boyd and Graham tried to standardize procedure of spasticity assessment and provided the modified TS (kept the 6-point system). MTS quantifies two angles: angle of ‘catch’ (stretch reflex threshold, measured at high speed) and the full range of motion (ROM, an angle measured at low speed). The difference between the two angles is suggested to be an indication of dynamic spasticity (dynamic tone).

Table 1 compares clinical evaluation and scores of spasticity by Ashworth and modified Ashworth scales. Table 2 demonstrates the method of clinical evaluation of spasticity by both Tardieu and modified Tardieu scales. The differences are only in performing the test and interpretation of the results. For passive stretch in MTS only two speeds are utilized. At slow speed, passive range of motion is found as an angle; and at fast speed, spastic reaction is evaluated as another angle. Tardieu et al. in 1954 experimented on passive moment of elbow at different speeds and had concluded that a stretch reflex was elicited at a specific speed (and above that) for each respective patient. The difference between these two angles is the true measure of spasticity for each patient.

Although clear protocols are used to teach and implement clinical scales, when a clinician repeats a scale or two clinicians implement the same scale on assessing a person’s level of spasticity, the results may not be the exact same. Objective and quantitative assessment methods make evaluation independent of individual judgement and often increase the sensitivity. They resolve the issues of inter- and at-rater variability as well. From objective and quantitative methods of spasticity assessment, in this review, we focus on one of the most popular biomechanical methods which is pendulum test.

Pendulum test

Introduced in 1951 Wartenberg’s pendulum test was designed to objectively assess knee spasticity in a free oscillation when the leg is first raised to horizontal position and then dropped [15]. This test is subjective, simple and quick to implement, reproducible, non-invasive, and is non-intimidating to the children or individuals with cognitive impairments. However, it has its own shortcomings. For example, the results are completely affected by the level of relaxation and by the form of sitting. This test was initially aimed explicitly at the lower limb, but modified versions were later used for upper extremities as well. Boczko et al. [27] modified it in 1958 as an instrumented test with a flashlight and camera to capture the trace of oscillation. Later, many sensor technologies have been used in instrumented pendulum test (Table 4). The common procedure to perform this test starts with the participant sitting in a relaxed position and the leg to test hanging freely over the edge of a seat or a table (Fig. 2). Then the examiner lifts the leg to a horizontal position and lets it swing freely. From this free oscillation, many measures can be extracted to quantify the level of spasticity. Since the test accuracy depends completely on the relaxation of the leg being tested (no voluntary muscle contraction), methods of confirmation of relaxation have been proposed. The two common methods are using Electromyography (EMG) [6, 9], and using phase plane diagram of angle versus angular velocity [28]. Numerous studies implemented this biomechanical test in diverse groups of patients including normal and spastic adults [10], healthy people and those in vegetative state [18], and on children with cerebral palsy [29]. All the results showed a good test–retest and inter-rater reliability of measurement, and that any of the used sensors can be reliably used to characterize leg kinematics during the Wartenberg pendulum test and provide quantitative evaluation of spasticity.

Fig. 2
figure 2

Experimental setup for pendulum test of spasticity at the knee level

To the best of our knowledge, there is no a comprehensive review that compares hardware or technologies for this test and also lists all possible outcome measures. So in this review, we have tried to provide a complete list of all possible and clinic-friendly technologies for implementation of this test along with their main advantages and drawbacks. We have also tried to come up with a comprehensive list of outcome measures.

Methods of implementation of pendulum test

Since its introduction in 1951, the pendulum test was implemented using various devices or technologies, and in different groups of patients with spasticity. However, it took 7 years until the first objective data collection was performed by Boczko and Mumenthaler using a battery-fed light and a video camera [30]. Most of the studies (93%) have focused on the spasticity at the knee joint, and a few tried to examine spasticity with pendulum test at elbow or ankle joints. In more than half a century, researchers have used this test in objective assessment of spasticity in many diseases including Stroke, Multiple Sclerosis (MS), Cerebral Palsy (CP), Spinal Cord Injury (SCI), lesions (cerebral or in central nervous system), paraplegia and hemiplegia, and trauma. Some of the studies (54%) compared healthy participants with their patient population. Almost half of these studies also used clinical scales either to choose the range of spasticity in their patient population [17, 31] or to compare their outcome measures according to the divisions that clinical scales provided [6, 18]. Table 3 presents available studies that used pendulum test in chronological order, indicating the joint under study, participant groups, main technology for implementation of the test, and whether clinical scales are used in the study. There have been important theoretical studies such that Jikuya et al. [32] that have modeled spasticity during the pendulum test, or Kusuhara et al. [30] and Yamamoto et al. [33] that have theoretically examined advantages of using two linear accelerometers in pendulum test, but not included in the table because of no experimental data on patient population.

Table 3 Summary of the studies and technological implementation of Wartenberg’s pendulum test

Discussion

Technologies used in objective pendulum test

Since the method of assessment of spasticity through pendulum test, and the processing steps to reach to the outcome measures are fully dependent on the implemented technology, this aspect is separately assessed in the literature. As Table 3 shows, variety of technologies, from ultrasound sensors [34] to camera-based [29, 35] or fiber-optic goniometers [36], have been used in the devices that implemented the test. However, these technologies differ in terms of price, stability, repeatability, ease of setup, complexity of signal post-processing, etc. advantages and disadvantages are compared in Table 4.

Table 4 Summary of the popular technologies used in objective pendulum test, and their advantages and shortcomings

Outcome measures of pendulum test

Alongside development of technologies of implementation for pendulum test, numerous measures were suggested to classify the severity of spasticity as outcome measures. Wartenberg used the test qualitatively paying attention to irregularities in swinging time or the number of oscillations. He also suggested Resting position as his method’s outcome measure. After that, many measures were introduced by each research group. A main goal of this review was to standardize the names and definitions that are given to the same aspect of pendulum-like oscillation (outcome measures) which does not look the same and is a source of confusion. This was done by grouping the all outcome measures based on the collected signals which are angle, angular velocity, and angular acceleration. Then, every possible category of suggested outcome measure, with all different names and definitions are brought under that group. This was performed in Table 5.

Table 5 Main categories of outcome measures (parameters) that are suggested in the studies that used pendulum test

The first mentioned set of outcome measures from pendulum test of spasticity on the knee were by Boczko et al. [27] in 1950s. During the following half a century, other than number of oscillations, test duration and peak angles, measures such as relaxation index (RI) were introduced to have higher reliability. The most comprehensive works on outcome measures during this time were performed by Bajd et al. [37,38,39] and Stillman et al. [19]. Table 5 illustrates all categories of outcome measures that have been suggested in the literature of pendulum test of spasticity. The three main categories are parameters based on angle data, based on angular velocity, and finally based on angular acceleration. All the investigated studies have used outcome measures that are based on joint angle during the test. Almost half of the studies have used parameters that are based on angular data (Table 5). A considerable amount of the studies has used more than one category (Table 5).

Overall, the suggested parameters can be divided into two categories of primary and secondary measures. Primary measures are directly obtained from aspects of the collected data such as the peak value of an angle, or number of oscillations (for example the first six categories in Table 5 for parameters based on angle data). Secondary outcome measures are calculated or combined from the primary ones mainly to reduce the number of parameters and to come up with ones that are more sensitive or better illustrate level of spasticity.

Outcome measures that are based on angle data are divided into 9 categories in Table 5. Category #1 is based on the initial position of the extended knee in the test. Although 29% of studies (out of 17 that we have investigated) have used this category in their set of parameters, it is not an outcome measure of spasticity test by itself, but is mainly used in the calculation of other outcome measures. Second category is used by 71% of the studies and relies on the maximum flexion angle in the first swing after the leg release. Category #3 examines the max angle in the first rebound towards extension of the knee, which is used by 18% of the studies. Resting angle or the final position of the leg is category #4 and is used by 41% of the literature. Category #5 examines durations in the test, whether overall duration or the time between two specific points, and is used by 35% of the studies. Number of cycles or frequency is category #6 and is used by 59% investigators. In counting the oscillations, a threshold of three degrees is considered below which the oscillations are ignored. Frequency concept was either used as the inverse of the time duration for the first cycle, or as the ratio of the number of total cycles to the total time. Time and frequency categories could be based on angular velocity or acceleration during the test, but almost all the researchers have based them on the angle data. Categories #7–9 are secondary measures and are calculated from the previous categories. Category #7 examines the ratio of the amplitudes for various angles. Wide range of ratios have been suggested by researchers and almost all studies (94%) have used at least one such category of outcome measure. Most of the powerful and consistent outcome measure that were capable of identifying levels of spasticity were in this category [40, 41]. Category #8 uses area under the knee angle curve which considers positive and negative values for flexion and extension around the resting angle, and is used by 24% of the studies. Recently, Popovic et al. suggested a total score for pendulum test which takes many aspects of the angle cure into account and comes up with a final score [35]. This is category #9 in Table 5.

There are three categories of outcome measures in the angular velocity group. Category #1 uses the first maximum angular velocity in the flexion direction, and is used by 35% of the investigators. Category #2 relies on the average angular velocity in the first swing towards the flexion, and is used by 12% of the literature. Category #3 is based on the first maximum velocity in the rebound towards the knee extension, and is suggested by the 18% of the literature. Some of the researchers call this category the minimum velocity instead of maximum velocity in the opposite direction.

Few studies have used outcome metrics that rely on angular acceleration. Stillman et al. [19] suggested the ratio of maximum acceleration to maximum deceleration during the first swing towards flexion of the knee as their measure of spasticity. Maximum angular acceleration both in flexion and in extension phases are separately used as outcome measures by Brown et al. [28] in 1988.

As can be seen from the table, quite a lot of parameters have been suggested as outcome measures. However, it is not practical to use all of the suggested parameters, and unfortunately no study has compared all these parameters to provide a comprehensive subjective scale of strength/weakness points. Although in Table 5 we provided an exhaustive list of outcome measures, in separated categories of measure signals, in Table 6 we introduce list of key publications that have used more than one outcome measure. In this way, pioneers in using popular sets of outcome measures are highlighted. Table 6 is the list of the main studies on the sets of outcome measures of pendulum test that is used in this review.

Table 6 Sets of outcome measures that were proposed in each key study

In the literature of the pendulum test for spasticity, there is no standard agreement between the researchers particularly in definition of the outcome measures. One main reason for this ambiguity is the different references used for measuring the knee angle. Overall, four key positions can be used as a reference: Horizontal plane, vertical plane, initial shank position, and final shank position in the test. Figure 3 shows the difference in the collected data with different experimental methodologies. Figure 3a shows these four possible frames of reference for knee angle, along the first four categories of angle parameters in Table 5.

Fig. 3
figure 3

a First couple of swing traces along the first four category of outcome measures (#1 to #4) that are key joint angles during the pendulum test and explained in Table 5. b Time series for the knee joint angle during the pendulum test along the first four category of outcome measures. The bold trace corresponds to when the reference is the resting angle. The dotted trace corresponds to when the reference is horizontal plane

Most of the studies have used horizontal plane as their frame of reference [6, 17,18,19, 28, 31, 42, 43]. The second most popular frame of reference for angular movement is rest angle or the final position of the knee in the pendulum test [29, 34, 35, 38, 44, 45]. Some of the researchers have used the initial angle of the shank as their reference instead of horizontal plane [36, 40, 46]. No researcher has used vertical plane as their reference in knee angle measurement. One example of such ambiguity is A0 which depending on the reference is defined as “Final resting angle” [40] as well as “knee angle at the beginning of the test during maximal limb extension” [18, 44]. Similarly, differences can be observed in the definition of other spasticity measures such is definition of RI in [6, 19] and in [38, 45].

Figure 3b illustrates two typical knee angle traces during a pendulum test. The bold trace corresponds to when the reference is the resting angle. Sections of knee angle trace, that parameters (angle outcome measures) in categories #1 to #4 are extracted, are marked. The dotted trace corresponds to when the reference is horizontal plane. If the beginning of trace falls on the time axis, then the reference is initial angle.

Conclusions

This systematic review aimed at analyzing the state of the art in implementation of Pendulum test and the outcome measures of this method of evaluation of spasticity. This test is agreed to be repeatable and a valid method of identifying presence of spasticity [29, 46], though it may not be successful in discriminating between close levels of spasticity [36]. Since lower leg weight is suitable to create enough speed through free fall in the pendulum test, which is crucial for this speed dependent symptom, most of this test’s applications is on the knee joint. However, using extra weight and special apparatus, this test can be applied to other joints such as elbow [47].

Non-instrumented pendulum test lacks objectivity and is apparently not used much by researchers. Instrumented pendulum test using multitudes of sensor technologies have been reported in the literature and evaluated as a feasible method of objective assessment of spasticity particularly for the knee extensors [19, 48, 49].

Sensor technologies that are used for implementation of this test (Tables 3, 4) would affect reliability/repeatability, ease and speed of setup, size and hindrance to pendular movement and hence suitability for clinical use, susceptibility to interfering factors such as gravity, ease of data processing and analysis and hence possibility of implementation in stand-alone devices. Overall, in the past few decades, the most technology been used is goniometer and electro-goniometer. Choice of sensor affects the type of signal that is used in data collection and thereby in outcome measure for the test. Although, processing techniques such as differentiation can transform angle data to angular velocity or acceleration. As can be seen in Table 5, all of the studies that we found to provide objective assessment method of spasticity with outcome metrics, used angular data. Almost half of them have also used angular velocity whether using tachometer/gyroscopes or differentiating angular data. Finally, almost 10% of the studies have focused on angular acceleration data as well.

We have categorized all 33 outcome measures that appeared in the literature into 13 categories in Table 5. These measures/metrics are supposed to help pendulum test differentiate between healthy and various levels of spasticity in the joint under consideration. A recent study also showed that the history of movement also affects the results of pendulum test [50]. Some of the studies have commented on the sensitivity of single outcome measures [6, 36, 51]. Each of the studies we investigated utilized between one to 14 outcome measures (Table 6).

Most of the studies have used more than one measure and argue that no single measure can represent all abnormal aspects of passive resistance of the disordered joint to the movement [19]. Unfortunately, there has been no study that optimizes a model using all or some of these 33 measures and comes up with a single combined measure with the highest discriminative power. Although such a model might be different for various spastic groups of patients. Nonetheless, some of the studies have tried such modeling approach without doing the optimization step. For example, Whelan et al. tried two models to find presence of knee spasticity in four different patient cohorts [36]. Another example is the attempt by popovic-Maneski et al. [35] in providing a single measure called global measure of spasticity.

Finally, such metrics that come out of pendulum test is supposed to be used by clinicians and should prove its power in a comparison by their golden clinical measures. Out of 5 clinical scales that are mentioned in studies we investigated, MAS and AS were the mostly used ones (Table 3). However, a modeling study that provides a relationship between such clinical scale as a gold standard and a single/combined outcome measure is really missing.

In this review we tried to come up with a comprehensive list of outcome measures along with the suggested sets of measures by lead researchers in this field (Tables 5, 6). We tried to provide standardized definitions for outcome measures through 13 categories for three type of signals (Table 5). However, it seems that a thorough modeling study is still missing. One that tries to take all of the 33 outcome measures into account and provides the best minimal sets of necessary parameters. Such final model, if provided, will accelerate making acceptable clinical device for objective assessment of spasticity. It goes without saying that there should be a simulation study on a mechanical pendulum with adjustable parameters, or alternatively a clinical experiment on patients with the whole range of spasticity before such optimization study can be performed.

Methods

Search strategy

A systematic review was conducted to identify literature published in four international databases between 1950 and the end of 2019: PubMed, Science Direct, Google Scholar and IEEE explore. Science Direct database was available from 1995 and hence was searched from 1995 to 2019. Combining the keywords “pendulum test” and “Spasticity” is defined as the search strategy. No filters were applied on the type of disease, because it was not the focus of study. The search strategy was the same in all databases, with modifications to fit the web interfaces. In other words, writing the queries or combination of Boolean operators could be modestly different for each database. Figure 1 shows the search strategy and selection process.

Study selection and eligibility criteria

After excluding duplicates, the selection of studies was carried out in two stages. All the remained articles were screened first by title and abstract, and then by full text for eligibility. At least one of the mentioned sections should include both “pendulum test” and “spasticity”, and also there should be a mention of quantitative/objective assessment, or instrument/sensor, or outcome measure. In other words, pure clinical studies with no quantitative evaluation of spasticity were excluded.

Availability of data and materials

The generated or analyzed data during this study are all included in the published article.

Abbreviations

CSI:

Composite Spasticity Index

HAT:

Hypertonia Assessment Tool

TSS:

Triple Spasticity Scale

AS:

Ashworth Scale

MAS:

Modified Ashworth Scale

TS:

Tardieu Scale

MTS:

Modified Tardieu Scale

ROM:

Range Of Motion

MS:

Multiple Sclerosis

CP:

Cerebral Palsy

SCI:

Spinal Cord Injury

RI:

Relaxation Index

NC:

No Comparison

DS:

Degenerative Syndromes of the nervous system

NR:

Not report in original paper

PTD:

Pendular Test Device

DAROM:

Dynamic Evaluation of Range of Motion scale

AIS:

Association Impairment Scale

ARSACS:

Autosomal Recessive Spastic Ataxia of Charlevoix/Saguenay

IMU:

Inertial measurement Units

ABI:

Acquired Brain Injury

SDR:

Spastic Diplegia undergoing selective dorsal Rhizotomy

Ang:

Angle

AMP:

Amplitude

Ex:

Excursion

TFR:

Time of first reversal

F, Freq:

Frequency

N:

Number of oscillation

Ncycle:

Number of cycle

ERI:

Extension Relaxation Index

AUC:

Area Under Curve

Vel:

Velocity

PT:

Total Pendulum Score

RA:

Rest Angle

EMG:

Electromyography

CNS:

Central Nervous System

References

  1. Pandyan A, Gregoric M, Barnes M, Wood D, Van WF, Burridge J, et al. Spasticity: Clinical perceptions, neurological realities and meaningful measurement. Disabil Rehabil. 2005;27:2–6. https://0-doi-org.brum.beds.ac.uk/10.1080/09638280400014576.

    Article  Google Scholar 

  2. Simpson DM, Gracies JM, Yablon SA, Barbano R, Brashear A. Botulinum neurotoxin versus tizanidine in upper limb spasticity: a placebo-controlled study. J Neurol Neurosurg Psychiatry. 2008;80:380–5. https://0-doi-org.brum.beds.ac.uk/10.1136/jnnp.2008.159657.

    Article  Google Scholar 

  3. Kirshblum S. Treatment alternatives for spinal cord injury related spasticity. J Spinal Cord Med. 1999;22:199–217. https://0-doi-org.brum.beds.ac.uk/10.1080/10790268.1999.11719570.

    Article  Google Scholar 

  4. Sheean G. The pathophysiology of spasticity. Eur J Neurol. 2002;9(Suppl 1):3–9. https://0-doi-org.brum.beds.ac.uk/10.1088/1751-8113/44/8/085201.

    Article  Google Scholar 

  5. Mayer NH. Clinicophysiologic concepts of spasticity and motor dysfunction in adults with an upper motoneuron lesion. Muscle Nerve Suppl. 1997;6:S1-13.

    Google Scholar 

  6. Greenan Fowler E, Nwigwe AI, Wong HT. Sensitivity of the pendulum test for assessing spasticity in persons with cerebral palsy. Dev Med Child Neurol. 2000;42:S0012162200000323. https://0-doi-org.brum.beds.ac.uk/10.1017/S0012162200000323.

    Article  Google Scholar 

  7. Burke D. Spasticity as an adaptation to pyramidal tract injury. Adv Neurol. 1988;47:401–23.

    Google Scholar 

  8. Jonnalagedda P, Deng F, Douglas K, Chukoskie L, Yip M, Ng TN, et al. An instrumented glove for improving spasticity assessment. In: 2016 IEEE Healthcare Innovation Point-Of-Care Technologies Conference (HI-POCT). New York: IEEE; 2016. p. 167–70. doi:10.1109/HIC.2016.7797723

  9. Calota A, Feldman AG, Levin MF. Spasticity measurement based on tonic stretch reflex threshold in stroke using a portable device. Clin Neurophysiol. 2008;119:2329–37.

    Article  Google Scholar 

  10. Decq P, Filipetti P, Lefaucheur JP. Evaluation of spasticity in adults. Oper Tech Neurosurg. 2004;7(3):100–8.

    Article  Google Scholar 

  11. Ashworth B. Preliminary trial of carisoprodol in multiple sclerosis. Practitioner. 1964;192:540–2.

    Google Scholar 

  12. Bohannon R, Smith M. Interrater reliability of a modified ashworth scale of muscle spasticity. Phys Ther. 1987;67:206–7.

    Article  Google Scholar 

  13. Tardieu G, Shentoub S, Delarue R. Research on a technic for measurement of spasticity. Rev Neurol. 1954;91:143.

    Google Scholar 

  14. Boyd RN, Graham HK. Objective measurement of clinical findings in the use of botulinum toxin type A for the management of children with cerebral palsy. Eur J Neurol. 1999;6(Suppl. 4):s23-35. https://0-doi-org.brum.beds.ac.uk/10.1111/j.1468-1331.1999.tb00031.x.

    Article  Google Scholar 

  15. Wartenberg R. Pendulousness of the legs as a diagnostic test. Neurology. 1951;1:18–18. https://0-doi-org.brum.beds.ac.uk/10.1212/wnl.1.1.18.

    Article  Google Scholar 

  16. Freddolini M, Corvi A, Barni L, Esposito F. Data processing techniques may influence numerical results and interpretation of single leg stance test. IRBM. 2020.

  17. Nordmark E, Andersson G. Wartenberg pendulum test: Objective quantification of muscle tone in children with spastic diplegia undergoing selective dorsal rhizotomy. Dev Med Child Neurol. 2002;44:26–33. https://0-doi-org.brum.beds.ac.uk/10.1111/j.1469-8749.2002.tb00255.x.

    Article  Google Scholar 

  18. Sterpi I, Caroli A, Meazza E, Maggioni G, Pistarini C, Colombo R. Lower limb spasticity assessment using an inertial sensor: a reliability study. Physiol Meas. 2013;34:1423–34. https://0-doi-org.brum.beds.ac.uk/10.1088/0967-3334/34/11/1423.

    Article  Google Scholar 

  19. Stillman B, McMeeken J. A video-based version of the pendulum test: technique and normal response. Arch Phys Med Rehabil. 1995;76:166–76.

    Article  Google Scholar 

  20. Ferreira DM, Liang H, Wu J. Knee joint kinematics of the pendulum test in children with and without Down syndrome. Gait Posture. 2020;76:311–7.

    Article  Google Scholar 

  21. Flamand V, Massé-Alarie H, Schneider C. Psychometric evidence of spasticity measurement tools in cerebral palsy children and adolescents: a systematic review. J Rehabil Med. 2013;45:14–23. https://0-doi-org.brum.beds.ac.uk/10.2340/16501977-1082.

    Article  Google Scholar 

  22. Bar-On L, Aertbeliën E, Molenaers G, Dan B, Desloovere K. Manually controlled instrumented spasticity assessments: a systematic review of psychometric properties. Dev Med Child Neurol. 2014;56:932–50. https://0-doi-org.brum.beds.ac.uk/10.1111/dmcn.12419.

    Article  Google Scholar 

  23. Biering-Sørensen F, Nielsen JB, Klinge K. Spasticity-assessment: a review. Spinal Cord. 2006;44:708–22.

    Article  Google Scholar 

  24. Poon DMY, Hui-Chan CWY. Hyperactive stretch reflexes, co-contraction, and muscle weakness in children with cerebral palsy. Dev Med Child Neurol. 2009;51:128–35.

    Article  Google Scholar 

  25. Albright L, Andrews M. Development of the Hypertonia Assessment Tool (HAT). Dev Med Child Neurol. 2009;52:411–2. https://0-doi-org.brum.beds.ac.uk/10.1111/j.1469-8749.2009.03477.x.

    Article  Google Scholar 

  26. Li F, Wu Y, Xiong L. Reliability of a new scale for measurement of spasticity in stroke patients. J Rehabil Med. 2014;46:746–53.

    Article  Google Scholar 

  27. Boczko M, Mumenthaler M. Modified pendulousness test to assess tonus of thigh muscles in spasticity. Neurology. 1958;8:846–846. https://0-doi-org.brum.beds.ac.uk/10.1212/wnl.8.11.846.

    Article  Google Scholar 

  28. Brown RA, Lawson DA, Leslie GC, Part NJ. Observations on the applicability of the Wartenberg pendulum test to healthy, elderly subjects. J Neurol Neurosurg Psychiatry. 1988;51:1171–7. https://0-doi-org.brum.beds.ac.uk/10.1136/jnnp.51.9.1171.

    Article  Google Scholar 

  29. Syczewska M, Lebiedowska MK, Pandyan AD. Quantifying repeatability of the Wartenberg pendulum test parameters in children with spasticity. J Neurosci Methods. 2009;178:340–4. https://0-doi-org.brum.beds.ac.uk/10.1016/j.jneumeth.2008.12.031.

    Article  Google Scholar 

  30. Kusuhara T, Jikuya K, Nakamura T, Michinishi H, Yamamoto Y, Okamoto T. Detection of knee-joint motions using two linear accelerometers for pendulum test. Electron Commun Japan. 2012;95:41–8. https://0-doi-org.brum.beds.ac.uk/10.1002/ecj.11383.

    Article  Google Scholar 

  31. Tancredo JR, Maria RM, de Azevedo ERFBM, Alonso KC, Varoto R, Cliquet JA. Clinical assessment of spasticity in individuals with spinal cord injury. Acta Ortopédica Bras. 2013;21:310–4. https://0-doi-org.brum.beds.ac.uk/10.1590/S1413-78522013000600002.

    Article  Google Scholar 

  32. Jikuya K, Michinishi H, Nakamura T, Kusuhara T, Yamamoto Y, Okamoto T. A detailed model for knee joint acceleration motions obtained by pendulum test. IFMBE Proc. 2009;25:472–5.

    Article  Google Scholar 

  33. Yamamoto Y, Jikuya K, Kusuhara T, Nakamura T, Michinishi H, Okamoto T. Precise measurement system for knee joint motion during the pendulum test using two linear accelerometers. In: Advanced Topics in Measurements. London: InTech; 2012. p. 19–42.

  34. Kaeser HE, Gihring H, Bergmann R, Gfhring I, Wachter KC, Ettlin TM. Testing an antispasticity drug (Tetrazepam) with the pendulum test: a monocentric pilot study. Neurorehabil Neural Repair. 1998;12:169–77. https://0-doi-org.brum.beds.ac.uk/10.1177/154596839801200405.

    Article  Google Scholar 

  35. Popovic-Maneski L, Aleksic A, Metani A, Bergeron V, Cobeljic R, Popovic DB. Assessment of spasticity by a pendulum test in SCI patients who exercise FES cycling or receive only conventional therapy. IEEE Trans Neural Syst Rehabil Eng. 2018;26:181–7.

    Article  Google Scholar 

  36. Whelan A, Sexton A, Jones M, O’Connell C, McGibbon CA. Predictive value of the pendulum test for assessing knee extensor spasticity. J Neuroeng Rehabil. 2018;15:68. https://0-doi-org.brum.beds.ac.uk/10.1186/s12984-018-0411-x.

    Article  Google Scholar 

  37. Popović-Maneski L, Aleksić A, Čobeljić R, Bajd T, Popović DB. A new method and instrumentation for analyzing spasticity. IETI Trans Ergon Saf. 2017;1:12–27. https://ieti.net/TES/2017V1I103.html.

  38. Bajd T, Vodovnik L. Pendulum testing of spasticity. J Biomed Eng. 1984;6:9–16.

    Article  Google Scholar 

  39. Aleksić A, Graovac S, Maneski LP, Popović DB. The assessment of spasticity: pendulum test based smart phone movie of passive markers. Serbian J Electr Eng. 2018;15:29–39.

    Article  Google Scholar 

  40. Bui HT, Gagnon C, Audet O, Mathieu J, Leone M. Measurement properties of a new wireless electrogoniometer for quantifying spasticity during the pendulum test in ARSACS patients. J Neurol Sci. 2017;375:181–5. https://0-doi-org.brum.beds.ac.uk/10.1016/j.jns.2017.01.065.

    Article  Google Scholar 

  41. Vargas Luna JL, Guðfinnsdóttir HK, Magnúsdóttir G, Guðmundsdóttir V, Krenn M, Mayr W, et al. Effects of sustained electrical stimulation on spasticity assessed by the pendulum test. Curr Dir Biomed Eng. 2016;2:405–7. https://0-doi-org.brum.beds.ac.uk/10.1515/cdbme-2016-0090.

    Article  Google Scholar 

  42. Leslie G, Muir C, Part N, Roberts R. A comparison of the assessment of spasticity by the Wartenberg pendulum test and the Ashworth grading scale in patients with multiple sclerosis. Clin Rehabil. 1992;6:41–8. https://0-doi-org.brum.beds.ac.uk/10.1177/026921559200600106.

    Article  Google Scholar 

  43. Szopa A, Domagalska-Szopa M, Kidoń Z, Syczewska M. Quadriceps femoris spasticity in children with cerebral palsy: measurement with the pendulum test and relationship with gait abnormalities. J Neuroeng Rehabil. 2014;11:166. https://0-doi-org.brum.beds.ac.uk/10.1186/1743-0003-11-166.

    Article  Google Scholar 

  44. He J, Norling WR, Wang Y. A dynamic neuromuscular model for describing the pendulum test of spasticity. IEEE Trans Biomed Eng. 1997;44:175–84. https://0-doi-org.brum.beds.ac.uk/10.1109/10.554764.

    Article  Google Scholar 

  45. Yeh C-H, Hung C-Y, Wang Y-H, Hsu W-T, Chang Y-C, Yeh J-R, et al. Novel application of a Wii remote to measure spasticity with the pendulum test: Proof of concept. Gait Posture. 2016;43:70–5. https://0-doi-org.brum.beds.ac.uk/10.1016/j.gaitpost.2015.10.025.

    Article  Google Scholar 

  46. Bohannon RW, Harrison S, Kinsella-Shaw J. Reliability and validity of pendulum test measures of spasticity obtained with the Polhemus tracking system from patients with chronic stroke. J Neuroeng Rehabil. 2009;6:30. https://0-doi-org.brum.beds.ac.uk/10.1186/1743-0003-6-30.

    Article  Google Scholar 

  47. Lin C-C, Ju M-S, Lin C-W. The pendulum test for evaluating spasticity of the elbow joint. Arch Phys Med Rehabil. 2003;84:69–74. https://0-doi-org.brum.beds.ac.uk/10.1053/apmr.2003.50066.

    Article  Google Scholar 

  48. Jamshidi M, Smith AW. Clinical measurement of spasticity using the pendulum test: Comparison of electrogoniometric and videotape analyses. Arch Phys Med Rehabil. 1996;77:1129–32.

    Article  Google Scholar 

  49. Valle MS, Casabona A, Sgarlata R, Garozzo R, Vinci M, Cioni M. The pendulum test as a tool to evaluate passive knee stiffness and viscosity of patients with rheumatoid arthritis. BMC Musculoskelet Disord. 2006;7:89.

    Article  Google Scholar 

  50. Willaert J, Desloovere K, Van Campenhout A, Ting LH, De Groote F. Movement history influences pendulum test kinematics in children with spastic cerebral palsy. Front Bioeng Biotechnol. 2020;8:920.

    Article  Google Scholar 

  51. Bajd T, Bowman B. Testing and modelling of spasticity. J Biomed Eng. 1982;4:90–6.

    Article  Google Scholar 

  52. Couvée LMJ, van der Laarse WD, Oosterveld WJ. Clinical experiences on spasticity with a modification of the Mumenthaler Pendulum Test. Paraplegia. 1968;6:96–102. https://0-doi-org.brum.beds.ac.uk/10.1038/sc.1968.17.

    Article  Google Scholar 

  53. Le Cavorzin P, Hernot X, Bartier O, Allain H, Carrault G, Rochcongar P, et al. A computed model of the pendulum test of the leg for routine assessment of spasticity in man. ITBM-RBM. 2001;22:170–7. https://0-doi-org.brum.beds.ac.uk/10.1016/S1297-9562(01)90028-2.

    Article  Google Scholar 

  54. LeMoyne R, Mastroianni T, Grundfest W, Nishikawa K. Implementation of an iPhone wireless accelerometer application for the quantification of reflex response. In: 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC). New York, NY: IEEE; 2013. p. 4658–61.

  55. Azevedo E, Alonso K, Maria R, Beinotti F, Cliquet A. Posture influence on the pendulum test of spasticity in spinal cord injured patients. J Neurol Sci. 2013;333:e567. https://0-doi-org.brum.beds.ac.uk/10.1016/j.jns.2013.07.1987.

    Article  Google Scholar 

  56. Popovic D, Bajd T. Pendulum test: Quantified assessment of the type and level of spasticity in persons with central nervous system lesions. Serbian J Electr Eng. 2018;15:1–12. https://0-doi-org.brum.beds.ac.uk/10.2298/SJEE171204006P.

    Article  Google Scholar 

  57. Haq Z. Advanced Topics in Measurements. 2012.

  58. Lin Y-C, Lin I-L, Chou T-FA, Lee H-M. Quantitative evaluation for spasticity of calf muscle after botulinum toxin injection in patients with cerebral palsy: a pilot study. J Neuroeng Rehabil. 2016;13:25. https://0-doi-org.brum.beds.ac.uk/10.1186/s12984-016-0135-8.

    Article  Google Scholar 

  59. Choi S, Kim J. Improving modified tardieu scale assessment using inertial measurement unit with visual biofeedback. In: 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC). New York, NY: IEEE; 2016. p. 4703–6.

Download references

Acknowledgements

Not applicable.

Funding

No commercial company was involved in this study and no external sources of funding were used.

Author information

Authors and Affiliations

Authors

Contributions

FR and RE conceived the work and designed the article. RE and NS performed the literature review. FR and RE wrote and revised the manuscript. FR, MRA, and NS edited the manuscript. FR supervised the work. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Fariborz Rahimi.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Rahimi, F., Eyvazpour, R., Salahshour, N. et al. Objective assessment of spasticity by pendulum test: a systematic review on methods of implementation and outcome measures. BioMed Eng OnLine 19, 82 (2020). https://0-doi-org.brum.beds.ac.uk/10.1186/s12938-020-00826-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s12938-020-00826-8

Keywords