Skip to main content

Table 2 Results of marker accuracy following exposure to 3000 cardiac cycles for each algorithm.

From: Multi-component based cross correlation beat detection in electrocardiogram analysis

Marker

Tolerance SD ms

Non-Syntactic SD ms

CC SD ms

Multi-component based CC SD ms

P-onset

10.2

22.6

22.1

11.9

P amplitude

 

23.4

23.8

7.8

P-offset

12.7

16.7

19.7

11.6

QRS-onset

6.5

10.4

10.1

6.6

QRS amplitude

 

14.3

1.8

1.8

QRS-offset

11.6

12.8

13.1

6.9

T amplitude

 

19.2

21.4

8.2

T-offset

30.6

18.7

20.6

14.6