Figure 6 | BioMedical Engineering OnLine

Figure 6

From: Current density imaging sequence for monitoring current distribution during delivery of electric pulses in irreversible electroporation

Figure 6

Measured electric current during each of the applied electric pulses by means of current probe (marked with blue crosses) and reconstructed electric current by means of CDI (marked with red circles) during the application of IRE pulse sequence. The CDI values were obtained as surface integrals of current density of over the area encircling one of the electrodes. Only the first and every tenth pulse are shown for easier comparison (a). The increase of the amplitude during each pulse is presented with 6 measurements obtained by the current probe. Here, measurements obtained on the last (90th) pulse are shown (b).