Skip to main content

Table 3 Variations in the working load and LVR values in the LV before and during counter-pulsation control

From: Effect of counter-pulsation control of a pulsatile left ventricular assist device on working load variations of the native heart

Animal

Load of LV (Joules/beat)

LVR of LV (%)

LV load reduction (%)

Before

During

Before

During

1

0.258 ± 0.113 (n = 115)

0.238 ± 0.055 (n = 113)

64.5 (n = 114)

9.8 (n = 112)

7.6

2

0.273 ± 0.106 (n = 130)

0.318 ± 0.034 (n = 130)

60.0 (n = 129)

10.4 (n = 129)

−16.5

3

0.216 ± 0.123 (n = 120)

0.148 ± 0.035 (n = 120)

73.7 (n = 119)

8.7 (n = 119)

31.5

  1. Data are presented as means ± standard deviation format.