Table 2 AED1 tests results

From: Radiated radiofrequency immunity testing of automated external defibrillators - modifications of applicable standards are needed

Device designation

Type of interference observed

Field strength for worst-case failure (V/m)

Modulation(Hz)

Frequency range of failure (MHz)

Device's behavior (Observations from user standpoint)

Notes

AED1

False negative

13

1 Hz pulsed (square wave)

39-44

Verbal response to fibrillation: "No shock advised"

V-Fib applied - no shock advised. Interference caused distorted ECG; analyzed as NSR. Analysis delayed during RF exposure.

AED1

False positive

20

4 Hz pulsed (square wave)

50 (single frequency test)

After analysis AED audio prompt:

"Shock advised" and "shock advised" appeared in the ECG printout

NSR applied but shock advised.

Interference masked the normal heart rhythm at a higher frequency