Table 1 Accuracy of the measurement setup depending on the number of cameras tracking a marker

From: 3D video-based deformation measurement of the pelvis bone under dynamic cyclic loading

Marker visible by n cameras 2 3 6 10 10 static
Noise level [+/- mm] at 0.01 mm step procedure 0.036 0.029 0.021 0.021 0.020
Noise level [+/- mm] at 0.1 mm step procedure 0.035 0.031 0.022 0.018 0.016
\